儿童腹部盆腔双能量与单能量CT的辐射剂量和图像质量比较:基于患者体型的研究

Comparison of radiation dose and image quality between contrast-enhanced single- and dual-energy abdominopelvic computed tomography in children as a function of patient size.
PubMed 文献 · Pediatr Radiol · 2021-07-09

本研究回顾性纳入860名儿童(12.3±5.3岁),按5种有效直径匹配为单能量CT(SECT)和双能量CT(DECT)各430例,比较腹部盆腔增强检查的辐射剂量与图像质量。结果显示DECT的SSDE和CTDIvol总体低于SECT,仅在最小直径组(<15 cm)CTDIvol与SECT相当;两组客观图像质量无显著差异。结论认为儿童腹部盆腔DECT可在相似或更低剂量下获得与SECT相似的图像质量,不受有效直径影响。

要点

  • 儿童腹部盆腔增强DECT的SSDE和CTDIvol总体低于SECT,图像质量相当。
  • 在有效直径<15 cm的小体型儿童中,DECT的CTDIvol与SECT无显著差异。
  • DECT可作为儿童腹部盆腔增强检查的替代方案,不增加辐射剂量。

核心结论

  • 在儿童腹部盆腔增强检查中,DECT的辐射剂量(SSDE和CTDIvol)总体低于或相当于SECT,且客观图像质量相似,不受患者有效直径影响。

研究设计与样本

  • 回顾性研究,纳入860名儿童(年龄12.3±5.3岁),按5种有效直径匹配为SECT组和DECT组各430例,均使用第二代双源CT行腹部盆腔增强检查,比较CTDIvol和SSDE及客观图像质量。

诊断标准与关键数值

  • DECT组平均SSDE为8.5±1.8 mGy,低于SECT组的9.3±2.0 mGy(P≤0.001)。
  • DECT组平均CTDIvol为5.6±2.4 mGy,低于SECT组的6.1±2.7 mGy(P≤0.001);但在有效直径<15 cm组中两组CTDIvol相当(P=0.065)。
  • 两组客观图像质量与有效直径的关系无显著差异(P>0.05)。

临床应用与阅片要点

  • 对于儿童腹部盆腔增强CT,可优先考虑DECT以降低辐射剂量,尤其适用于有效直径≥15 cm的患儿。
  • 在有效直径<15 cm的小体型儿童中,DECT的CTDIvol与SECT相当,仍需关注辐射剂量优化。
  • DECT在降低剂量的同时保持与SECT相似的客观图像质量,支持其临床推广应用。

局限性

  • 回顾性研究设计,可能存在选择偏倚。
  • 单中心研究,结果外推需谨慎。
  • 仅评估客观图像质量,未涉及主观图像质量或诊断性能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Widespread adoption of dual-energy computed tomography (DECT) requires evidence it does not cause higher radiation dose than conventional single-energy CT (SECT). While a few publications involving pediatric patients exist, most have focused on small cohorts. Hence, there is still a need for studies that ascertain what radiation doses are expected in larger populations that include representative ranges of patient sizes and ages.

OBJECTIVE: To compare radiation dose and image quality of DECT and SECT abdominopelvic examinations in children as a function of patient size.

MATERIALS AND METHODS: This retrospective study included 860 children (age range: 12.3±5.3 years) who underwent contrast-enhanced abdominopelvic exams on second-generation dual-source CT in a five-year period. Two groups, SECT and DECT, consisting of 430 children each, were matched by 5 effective diameters. Volume CT dose index (CTDIvol) and size-specific dose estimate (SSDE) were analyzed as a function of effective diameter. Objective image quality was compared between the groups.

RESULTS: DECT SSDEs were lower across all effective patient diameters compared with SECT (mean: 8.5±1.8 mGv vs. 9.3±2.0 mGv, respectively, P≤0.001). DECT CTDIvol was lower compared to SECT (mean: 5.6±2.4 mGv vs. 6.1±2.7 mGv, respectively, P≤0.001) except in the smallest diameter group (<15 cm) where it was comparable to SECT (P=0.065). Objective image quality versus effective diameter between the two CT groups was comparable (P>0.05).

CONCLUSION: In children, regardless of effective diameter, contrast-enhanced abdominopelvic DECT can be performed with a similar or lower dose and similar image quality compared with SECT examinations.

来源:Pediatr Radiol,查看原文。版权归原作者所有。