CAIPIRINHA加速3D脂肪饱和SPACE MRI与2D MRI评估肩部病变的对比
以2D TSE为参考标准,3D SPACE对肌腱异常、肩锁关节异常、邻近骨改变及关节积液/积液的敏感度、特异度和准确度均较高。结论认为6分钟3D SPACE与10分钟2D TSE性能相似,可在不降低图像质量的前提下缩短扫描时间。
要点
- 3D SPACE序列扫描时间仅6分钟,图像质量与10分钟2D TSE相似。
- 以2D TSE为参考,3D SPACE对肌腱异常敏感度94.81%、特异度94.12%。
- 3D SPACE对关节积液/积液评估敏感度97.30%、特异度100%。
关键影像表现/征象
- 3D图像噪声标准差低于2D图像,液体对比噪声比高于2D图像(p=0.00)。
- 图像质量评估中2D与3D CAIPIRINHA SPACE相似(p=0.05)。
- 方法间一致性为中等至几乎完美,观察者间一致性也为中等至几乎完美,其中SSP病变一致性最高。
诊断标准与关键数值
- 以2D TSE为参考标准,3D SPACE对肌腱异常的敏感度94.81%、特异度94.12%、准确度94.39%。
- 以2D TSE为参考标准,3D SPACE对肩锁关节异常的敏感度97.06%、特异度80.00%、准确度91.84%。
- 以2D TSE为参考标准,3D SPACE对邻近骨改变的敏感度88.89%、特异度100.00%、准确度93.89%。
- 以2D TSE为参考标准,3D SPACE对关节积液/积液评估的敏感度97.30%、特异度100%、准确度97.96%。
临床应用与阅片要点
- 3D脂肪饱和CAIPIRINHA SPACE可用于肩部MRI缩短扫描时间,图像质量不劣于2D TSE。
- 诊断肌腱异常、肩锁关节异常、骨改变及关节积液时,3D SPACE与2D TSE一致性中等至几乎完美,可支持日常报告。
- 肩锁关节异常特异度相对较低(80.00%),阅片时需注意结合其他征象。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To compare the performance of 6-min MRI with a fat-saturated 3D-controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) Sampling perfection with application-optimized contrast using different flip angle evolution (SPACE) TSE protocol with 10-min 2D TSE MRI protocol for assessment of abnormalities of the shoulder.
METHODS: Forty-nine subjects underwent both 3D fat-saturated-CAIPIRINHA SPACE and 2D TSE sequences of the shoulder on a 3.0-T system. Following randomization and anonymization, two musculoskeletal radiologists evaluated the 2D and 3D images independently for image quality and diagnostic capability. Descriptive statistics, inter-observer, and inter-method concordance were investigated. p values < 0.05 were considered significant.
RESULTS: For image quality assessment, 2D images were similar to 3D CAIPIRINHA SPACE images (p = 0.05). 3D had lower noise standard deviation (SD) and higher fluid CNR than 2D images (p = 0.00). For diagnostic capability assessment, using 2D TSE as a standard of reference, sensitivity, specificity, and accuracy of 3D SPACE were, respectively, 94.81%, 94.12%, and 94.39% for tendon abnormalities; 97.06%, 80.00%, and 91.84% for acromioclavicular joint abnormalities; 88.89%, 100.00%, and 93.89% for adjacent bone alterations; and 97.30%, 100%, and 97.96% for joint fluid/effusion assessment. The inter-method concordance was moderate to almost perfect. The inter-observer-concordance of the shoulder assessment was also moderate to almost perfect, with SSP lesions demonstrating the greatest concordance.
CONCLUSIONS: The performance of 6-min 3D fat-saturated-CAIPIRINHA SPACE MRI for shoulder MRI is similar to that of 10-min 2D TSE MRI. 3D fat-saturated-CAIPIRINHA SPACE MRI can be utilized to reduce scan time without degradation in image quality.
KEY POINTS: • CAIPIRINHA acceleration 3D fat-saturated-MRI of the shoulder is achievable in 6 min with high spatial resolution. • 3D fat-saturated CAIPIRINHA MRI is similar to 2D MRI in the shoulder assessment. • 3D CAIPIRINHA MRI images enable rapid diagnosis of shoulder abnormalities without image quality degradation.