双臂下垂位腹部CT:虚拟单能重建对诊断图像质量的影响
回顾性纳入162例双臂下垂行双层探测器CT腹部增强扫描患者,比较120-kVp图像与40-100 keV虚拟单能图像。60-100 keV虚拟单能图像伪影指数显著低于120-kVp,70-100 keV主观条纹伪影显著更轻,70-keV被两位独立读者选为最佳图像,其病灶检出AUC显著高于120-kVp。
要点
- 双臂下垂位腹部CT时,70-keV虚拟单能图像可有效减轻手臂所致条纹伪影并提高图像对比度。
- 70-keV虚拟单能图像对肝、脾、肾病灶检出的AUC显著高于常规120-kVp图像。
- 60-100 keV虚拟单能图像伪影指数显著低于120-kVp,70-100 keV主观伪影严重程度显著更低。
关键影像表现/征象
- 60-100 keV虚拟单能图像的伪影指数显著低于120-kVp图像(p < 0.001)。
- 70-100 keV虚拟单能图像的手臂所致条纹伪影主观严重程度显著低于120-kVp图像(p < 0.001)。
- 40-70 keV虚拟单能图像在伪影区和参考区测得的CNR优于120-kVp图像(p < 0.05)。
- 60-90 keV虚拟单能图像的整体图像质量优于120-kVp图像(p < 0.05)。
- 两位独立读者均选择70-keV虚拟单能图像为最佳图像。
诊断标准与关键数值
- 肝、脾、肾病灶检出ROC分析中,70-keV虚拟单能图像AUC显著高于120-kVp图像:读者1为0.962 vs 0.891(p = 0.011),读者2为0.981 vs 0.922(p = 0.012)。
临床应用与阅片要点
- 对无法上举手臂行腹部CT的患者,可优先重建70-keV虚拟单能图像以减轻手臂条纹伪影并改善病灶检出。
- 60-100 keV虚拟单能图像均可降低伪影指数,但70-keV在主观图像质量和诊断性能间表现最佳。
局限性
- 本研究为回顾性设计。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate effect of virtual monoenergetic reconstruction on contrast-enhanced abdominal CT obtained using a dual-layer detector CT (DLCT) in patients with arms-down positioning, in comparison with conventional 120-kVp image.
METHODS: A total of 162 consecutive patients who underwent abdominal CT scans using a DLCT with arms-down positioning were retrospectively enrolled. 120-kVp image and virtual monoenergetic images (VMIs) at 40-100 keV were compared for arm-induced artifact reduction capability and image quality through objective and subjective assessments. Diagnostic performance was compared between VMI with the best quality and 120-kVp image.
RESULTS: VMIs at 60-100 keV showed significantly lower (p < 0.001) artifact index than 120-kVp image. Subjective severity of arm-induced streak artifacts was significantly (p < 0.001) lower in VMIs at 70-100 keV than in 120-kVp image. CNRs measured in artifact and reference areas were superior (p < 0.05) in 40-70 keV to those in 120 kVp. Overall image quality was superior (p < 0.05) in 60-90 keV to 120 kVp. 70-keV VMI was chosen as the best image by two independent readers. In ROC analysis of lesion detection in the liver, spleen, and kidneys, the AUC was significantly higher for 70-keV VMI (reader 1, 0.962; p = 0.011; reader 2, 0.981; p = 0.012) than that for 120-kVp image (reader 1, 0.891; reader 2, 0.922).
CONCLUSION: The70-keV VMI from DLCT could improve abdominal CT images by reducing streak artifacts and increasing image contrast in patients who could not elevated their arms. The 70-keV VMI showed better diagnostic performance than the conventional 120 kVp image.
引用
- Bae K, Jeon KN, Kim JE, et al. Abdominal CT in patients with arms down positioning: Effect of virtual monoenergetic reconstruction on diagnostic image quality. Abdom Radiol (NY). 2021. PMID: 34215901. doi:10.1007/s00261-021-03195-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。