不同驱动功率幅度对儿童肝脏MR弹性成像肝硬度测量的影响
本研究回顾性纳入16例儿童患者,评估20%与56%驱动功率幅度下肝脏MR弹性成像的肝硬度测量一致性。结果显示两种功率下肝硬度ICC为0.857(95%CI 0.760-0.915),可靠性好,但ROI面积及最大值、最小值、标准差的一致性仅0.33-0.51。作者建议根据儿童肝脏大小优化驱动功率幅度。
要点
- 儿童肝脏MRE中,20%与56%驱动功率幅度测得的肝硬度一致性良好(ICC=0.857)。
- 不同功率幅度下ROI面积及肝硬度最大值、最小值、标准差的一致性较差(ICC 0.33-0.51)。
- 儿童肝脏MRE应根据肝脏大小优化驱动功率幅度,以提高测量可靠性。
关键影像表现/征象
- 20%功率下肝脏ROI中位面积为83.1 cm2(范围46.9-144.1 cm2),56%功率下为63.0 cm2(范围5.4-123.4 cm2)。
- 20%功率下肝硬度中位值为2.3 kPa(范围1.7-8.0 kPa),56%功率下为2.8 kPa(范围1.7-8.5 kPa)。
- 20%功率下肝硬度最大值和最小值分别为5.3 kPa和1.0 kPa,56%功率下分别为7.8 kPa和1.1 kPa。
- 20%功率下肝硬度标准差为0.6 kPa,56%功率下为1.0 kPa。
诊断标准与关键数值
- 两种功率幅度间肝硬度ICC为0.857(95%CI 0.760-0.915)。
- ROI面积及肝硬度最大值、最小值、标准差的ICC为0.33-0.51。
临床应用与阅片要点
- 儿童肝脏MRE测量肝硬度时,应关注驱动功率幅度对ROI面积及硬度离散度的影响,根据肝脏大小选择合适的功率幅度。
局限性
- 回顾性研究,样本量较小(16例),仅纳入单中心数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess how different driver power amplitudes affect the measurement of liver stiffness in pediatric liver magnetic resonance elastography (MRE).
METHODS: From January 2018 to May 2018, pediatric patients (≤ 18 years) who underwent liver MRE with 20% and 56% driver power amplitudes were included in this retrospective study. Region-of-interests (ROIs) were drawn on four stiffness maps to include the largest area of the liver parenchyma. Intraclass correlation coefficients (ICCs) were used to assess agreements for the area, mean, maximum, minimum and standard deviation of liver stiffness between the driver power amplitudes.
RESULTS: 128 MRE stiffness maps from 16 patients (M:F = 10:6, median 12.5 years old) were included. On MRE, median ROI areas of liver were 83.1 cm2 (range, 46.9-144.1 cm2) and 63.0 cm2 (range, 5.4-123.4 cm2) for the driver power amplitudes of 20% and 56%, respectively. Median liver stiffness values were 2.3 kPa (range, 1.7-8.0 kPa) and 2.8 kPa (range, 1.7-8.5 kPa). Maximum and minimum liver stiffness values were 5.3 kPa and 1.0 kPa for 20% and 7.8 kPa and 1.1 kPa for 56%. Standard deviation was 0.6 kPa for 20% and 1.0 kPa for 56%. ICC values between the two power amplitudes were 0.33-0.51 for the ROI area and the maximum, minimum and standard deviation values of liver stiffness. The ICC value for liver stiffness was 0.857 (95% confidence interval, 0.760-0.915).
CONCLUSION: Liver stiffness with two driver power amplitudes on MRE showed good reliability in pediatric patients. Driver power amplitudes need to be optimized according to the pediatric liver size.