基于人工智能的计算机辅助检测系统在不同低剂量水平双源光子计数探测器CT中评估肺结节的初步性能评价
在含14个肺结节(3-12 mm)的体模上,于光子计数探测器CT和常规能量积分探测器CT上以3种扫描模式及递减剂量(CTDIvol 1.79-0.31 mGy、IQ 30-5)成像,光子计数CT主观图像质量优于剂量匹配的能量积分CT(P<0.001),AI-CAD在两者间肺结节检测与体积测量结果总体相当,但低剂量下Q+和Q模式假阳性率高于能量积分CT,需注意假阳性发现。
要点
- 光子计数CT主观图像质量优于剂量匹配的能量积分CT(P<0.001),客观噪声相当或更低。
- AI-CAD在光子计数CT与剂量匹配能量积分CT间肺结节检测与体积测量结果相当(P=0.08-1.00)。
- 最低评估剂量水平(IQ5)下光子计数CT平均敏感度95%、剂量匹配能量积分CT 86%。
核心结论
- 光子计数探测器CT主观图像质量优于剂量匹配的能量积分探测器CT,而客观图像噪声相当或更低;全自动AI辅助肺结节检测与体积测量在光子计数CT上可行,但须注意假阳性发现。
研究设计与样本
- 体模研究:在光子计数探测器CT和常规能量积分探测器CT上,对含14个不同大小(3-12 mm)肺结节的胸部体模,以3种厂商特定扫描模式(QuantumPlus、Quantum、High Resolution)在递减匹配辐射剂量水平(CTDIvol 1.79-0.31 mGy、IQ 30-5)下成像;胸壁8个位置手动测量图像噪声,2名读者共识评价主观图像质量,采用商用AI-CAD系统进行结节检测和体积测量。
关键影像表现/征象
- 主观图像质量光子计数CT优于能量积分CT(P<0.001)。
- 客观图像噪声在Q+和Q模式间相似(P>0.05),HR模式优于能量积分CT(光子计数CT 55.8±11.7 HU vs 能量积分CT 74.8±5.4 HU;P=0.01)。
- HR模式在光子计数各模式中图像噪声值最低(P=0.01)。
诊断标准与关键数值
- AI-CAD在光子计数CT与剂量匹配能量积分CT间肺结节检测和体积测量结果相当(P=0.08-1.00)。
- 最低评估剂量水平(IQ5)下,光子计数CT平均敏感度95%,剂量匹配能量积分CT平均敏感度86%。
- HR模式在最低评估剂量水平(IQ5;CTDIvol 0.41 mGy)敏感度100%、假阳性率1。
- 低剂量水平(IQ10和IQ5)下,Q+和Q模式假阳性率高于能量积分CT。
临床应用与阅片要点
- 光子计数CT在主观图像质量上优于剂量匹配能量积分CT,客观噪声相当或更低。
- 全自动AI辅助结节检测与体积测量在光子计数CT上可行,但须注意假阳性发现。
- 低剂量水平(IQ10和IQ5)下Q+和Q模式的假阳性率高于能量积分CT。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The aim of this study was to evaluate the image quality (IQ) and performance of an artificial intelligence (AI)-based computer-aided detection (CAD) system in photon-counting detector computed tomography (PCD-CT) for pulmonary nodule evaluation at different low-dose levels.
MATERIALS AND METHODS: An anthropomorphic chest-phantom containing 14 pulmonary nodules of different sizes (range, 3-12 mm) was imaged on a PCD-CT and on a conventional energy-integrating detector CT (EID-CT). Scans were performed with each of the 3 vendor-specific scanning modes (QuantumPlus [Q+], Quantum [Q], and High Resolution [HR]) at decreasing matched radiation dose levels (volume computed tomography dose index ranging from 1.79 to 0.31 mGy) by adapting IQ levels from 30 to 5. Image noise was measured manually in the chest wall at 8 different locations. Subjective IQ was evaluated by 2 readers in consensus. Nodule detection and volumetry were performed using a commercially available AI-CAD system.
RESULTS: Subjective IQ was superior in PCD-CT compared with EID-CT (P < 0.001), and objective image noise was similar in the Q+ and Q-mode (P > 0.05) and superior in the HR-mode (PCD 55.8 ± 11.7 HU vs EID 74.8 ± 5.4 HU; P = 0.01). High resolution showed the lowest image noise values among PCD modes (P = 0.01). Overall, the AI-CAD system delivered comparable results for lung nodule detection and volumetry between PCD- and dose-matched EID-CT (P = 0.08-1.00), with a mean sensitivity of 95% for PCD-CT and of 86% for dose-matched EID-CT in the lowest evaluated dose level (IQ5). Q+ and Q-mode showed higher false-positive rates than EID-CT at lower-dose levels (IQ10 and IQ5). The HR-mode showed a sensitivity of 100% with a false-positive rate of 1 even at the lowest evaluated dose level (IQ5; CDTIvol, 0.41 mGy).
CONCLUSIONS: Photon-counting detector CT was superior to dose-matched EID-CT in subjective IQ while showing comparable to lower objective image noise. Fully automatized AI-aided nodule detection and volumetry are feasible in PCD-CT, but attention has to be paid to false-positive findings.
引用
- Jungblut L, Blüthgen C, Polacin M, et al. First Performance Evaluation of an Artificial Intelligence-Based Computer-Aided Detection System for Pulmonary Nodule Evaluation in Dual-Source Photon-Counting Detector CT at Different Low-Dose Levels. Invest Radiol. 2022. PMID: 34324462. doi:10.1097/RLI.0000000000000814
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。