超快波控并行成像(CAIPI)磁化准备快速梯度回波(MP-RAGE)与标准MP-RAGE在非镇静儿童中的比较:初步临床经验

Comparison of ultrafast wave-controlled aliasing in parallel imaging (CAIPI) magnetization-prepared rapid acquisition gradient echo (MP-RAGE) and standard MP-RAGE in non-sedated children: initial clinical experience.
PubMed 文献 · Pediatr Radiol · 2021-07-15

该研究纳入24例清醒儿童,比较标准MP-RAGE与超快wave-CAIPI MP-RAGE脑成像序列的可行性。结果显示wave-CAIPI MP-RAGE使用32通道线圈时扫描时间缩短77%、20通道线圈时缩短70%,但观察者在中央及外周噪声和 anatomical结构评估上更偏好标准MP-RAGE,两者在病变显示及病理强化方面无显著差异。

要点

  • 超快wave-CAIPI MP-RAGE在清醒儿童脑成像中可行,扫描时间显著缩短。
  • 病变显示、伪影及病理强化方面,wave-CAIPI与标准MP-RAGE无显著差异。
  • 中央及外周噪声、解剖结构评估上,观察者更偏好标准MP-RAGE。

核心结论

  • 超快wave-CAIPI MP-RAGE用于清醒儿童脑成像可行,可大幅缩短扫描时间,但主观图像噪声增加。

研究设计与样本

  • 初步临床经验研究,纳入24例患者(男14例,平均年龄11.5±4.5岁,范围1个月至17.8岁),每次MRI均包含标准MP-RAGE和超快wave-MP-RAGE序列,由两名独立小儿神经放射科医生采用预设5点量表评估。

关键影像表现/征象

  • wave-CAIPI MP-RAGE使用32通道线圈时扫描时间缩短77%,使用20通道线圈时缩短70%;病变显示、伪影及病理强化(包括实质、软脑膜和硬脑膜强化)在标准MP-RAGE与wave-CAIPI MP-RAGE间无显著差异(均P>0.05);中央噪声(P<0.001)、外周噪声(P<0.001)及解剖结构评估(P<0.001)方面,观察者更偏好标准MP-RAGE。

临床应用与阅片要点

  • 原文未据此给出推荐,仅报告超快wave-CAIPI MP-RAGE在清醒儿童中可行且缩短扫描时间,但主观噪声增加及解剖结构评估偏好标准序列。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Fast magnetic resonance imaging (MRI) sequences are advantageous in pediatric imaging as they can lessen child discomfort, decrease motion artifact and improve scanner availability.

OBJECTIVE: To evaluate the feasibility of an ultrafast wave-CAIPI (controlled aliasing in parallel imaging) MP-RAGE (magnetization-prepared rapid gradient echo) sequence for brain imaging of awake pediatric patients.

MATERIALS AND METHODS: Each MRI included a standard MP-RAGE sequence and an ultrafast wave-MP-RAGE sequence. Two neuroradiologists evaluated both sequences in terms of artifacts, noise, anatomical contrast and pathological contrast. A predefined 5-point scale was used by two independent pediatric neuroradiologists. A Wilcoxon signed-rank test was used to evaluate the difference between sequences for each variable.

RESULTS: Twenty-four patients (14 males; mean age: 11.5±4.5 years, range: 1 month to 17.8 years) were included. Wave-CAIPI MP-RAGE provided a 77% reduction in scan time using a 32-channel coil and a 70% reduction using a 20-channel coil. Visualization of the pathology, artifacts and pathological enhancement (including parenchymal, leptomeningeal and dural enhancement) was not significantly different between standard MP-RAGE and wave-CAIPI MP-RAGE (all P>0.05). For central (P<0.001) and peripheral (P<0.001) noise, and the evaluation of the anatomical structures (P<0.001), the observers favored standard MP-RAGE over wave-CAIPI MP-RAGE.

CONCLUSION: Ultrafast brain imaging with wave-CAIPI MP-RAGE is feasible in awake pediatric patients, providing a substantial reduction in scan time at a cost of subjectively increased image noise.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Tabari A, Conklin J, Figueiro Longo MG, et al. Comparison of ultrafast wave-controlled aliasing in parallel imaging (CAIPI) magnetization-prepared rapid acquisition gradient echo (MP-RAGE) and standard MP-RAGE in non-sedated children: initial clinical experience. Pediatr Radiol. 2021. PMID: 34268599. doi:10.1007/s00247-021-05117-5