比较有无金属伪影去除算法的多种多色与单色双能CT图像质量以评估全膝关节置换术后

Comparison of the Quality of Various Polychromatic and Monochromatic Dual-Energy CT Images with or without a Metal Artifact Reduction Algorithm to Evaluate Total Knee Arthroplasty.
PubMed 文献 · Korean J Radiol · 2021-05-26

纳入58例患者74次全膝关节置换术后双能CT检查,比较多色图像(权重因子0.4、0、-0.3)与单色图像(130、150、170、190 keV)加或不加iMAR共14组图像质量。应用iMAR的多色图像P-0.3i和P0.0i图像质量排名最高,低密度伪影体积最小见于P-0.3i、P0.0i及iMAR单色130 keV,高密度伪影体积最小见于P-0.3i。

要点

  • iMAR可提高多色和单色双能CT图像质量排名,P-0.3i和P0.0i排名最高。
  • 低密度伪影体积最小见于P-0.3i、P0.0i及iMAR单色130 keV。
  • 高密度伪影体积最小见于P-0.3i,CNR最佳见于权重因子0.4多色图像。

核心结论

  • 应用iMAR的多色图像P-0.3i和P0.0i相比其他图像组提供更好的图像质量和显著的金属伪影减少。

研究设计与样本

  • 纳入58例患者(男28例、女30例;平均年龄71.4岁,范围61-83岁)共74次全膝关节置换术后双能CT检查,图像组包括线性混合80 kVp与锡滤过140 kVp权重因子0.4、0、-0.3的多色图像和130、150、170、190 keV单色图像,加或不加iMAR共14组;两名读者对图像质量进行1至14级定性排名,并在25个无金属伪影影响的膝关节子集中定量测量高、低密度伪影体积及骨与脂肪组织间CNR。

关键影像表现/征象

  • 应用iMAR的多色图像P-0.3i和P0.0i获得最高图像质量排名评分,一名读者对两者中位数均为14,另一名读者分别为13和14(p < 0.001)。
  • 所有应用iMAR的图像序列排名评分均高于未应用iMAR的对应序列。
  • 低密度伪影体积最小见于P-0.3i、P0.0i及应用iMAR的单色130 keV图像。
  • 高密度伪影体积最小见于P-0.3i。
  • CNR最佳见于权重因子0.4的多色图像(加或不加iMAR),其次为权重因子0.0的多色图像(加或不加iMAR)。

诊断标准与关键数值

  • 图像质量定性排名采用1(最低质量)至14(最高质量)的评分标准。
  • 定量测量包括高密度和低密度伪影体积以及骨与脂肪组织间的CNR,在25个无金属伪影影响的膝关节子集中进行。

临床应用与阅片要点

  • 在评估全膝关节置换术后双能CT时,应用iMAR的多色图像P-0.3i和P0.0i可提供更好的图像质量并显著减少金属伪影。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To compare the quality of various polychromatic and monochromatic images with or without using an iterative metal artifact reduction algorithm (iMAR) obtained from a dual-energy computed tomography (CT) to evaluate total knee arthroplasty.

MATERIALS AND METHODS: We included 58 patients (28 male and 30 female; mean age [range], 71.4 [61-83] years) who underwent 74 knee examinations after total knee arthroplasty using dual-energy CT. CT image sets consisted of polychromatic image sets that linearly blended 80 kVp and tin-filtered 140 kVp using weighting factors of 0.4, 0, and -0.3, and monochromatic images at 130, 150, 170, and 190 keV. These image sets were obtained with and without applying iMAR, creating a total of 14 image sets. Two readers qualitatively ranked the image quality (1 [lowest quality] through 14 [highest quality]). Volumes of high- and low-density artifacts and contrast-to-noise ratios (CNRs) between the bone and fat tissue were quantitatively measured in a subset of 25 knees unaffected by metal artifacts.

RESULTS: iMAR-applied, polychromatic images using weighting factors of -0.3 and 0.0 (P-0.3i and P0.0i, respectively) showed the highest image-quality rank scores (median of 14 for both by one reader and 13 and 14, respectively, by the other reader; p < 0.001). All iMAR-applied image series showed higher rank scores than the iMAR-unapplied ones. The smallest volumes of low-density artifacts were found in P-0.3i, P0.0i, and iMAR-applied monochromatic images at 130 keV. The smallest volumes of high-density artifacts were noted in P-0.3i. The CNRs were best in polychromatic images using a weighting factor of 0.4 with or without iMAR application, followed by polychromatic images using a weighting factor of 0.0 with or without iMAR application.

CONCLUSION: Polychromatic images combined with iMAR application, P-0.3i and P0.0i, provided better image qualities and substantial metal artifact reduction compared with other image sets.