文献速递腹部泌尿生殖儿科影像MR

快速B1不敏感MR指纹技术用于肾脏定量成像

Rapid B1-Insensitive MR Fingerprinting for Quantitative Kidney Imaging.
PubMed 文献 · Radiology · 2021-06-08

该研究设计了一种15秒肾脏MR指纹(MRF)采集方案,在体模、10名健康成人志愿者和3名常染色体隐性多囊肾病(ARPKD)患儿中评估其可行性。结果显示MRF肾脏T1、T2值与文献一致且精度提高,B1校正与否结果高度一致(R2>0.96,P<.001),表明对B1场不均匀性敏感性有限,ARPKD患者重复扫描重现性好(T1和T2差异<3%)。

要点

  • 15秒肾脏MRF单次屏气即可获得T1、T2定量图,无需B1校正。
  • 健康成人肾皮质T1约1362毫秒、T2约64毫秒,髓质T1约1827毫秒、T2约69毫秒。
  • ARPKD患者连续日重复扫描T1和T2差异<3%,重现性良好。

核心结论

  • 肾脏MR指纹方法可在3.0T单次屏气下提供在体肾脏T1和T2图,精度提高且无需B1校正。

研究设计与样本

  • 研究在体模、健康志愿者和患者中评估快速肾脏MRF技术;采用15秒采集、12个采集段、低翻转角5°-12°、多种磁化准备方案及欠采样螺旋轨迹;2019年8月至2020年10月纳入10名健康成人志愿者(平均年龄35岁±13,6名男性)和3名ARPKD患儿(平均年龄10岁±3,2名男孩)。

关键影像表现/征象

  • 健康成人志愿者肾皮质T1为1362毫秒±5、T2为64毫秒±5,肾髓质T1为1827毫秒±94、T2为69毫秒±3。
  • 有无B1校正的MRF肾脏T1和T2值高度一致(R2>0.96,P<.001)。
  • 使用前9个采集段(11秒采集时间)与12个采集段(15秒采集时间)的重建结果高度一致(R2>0.87,P<.001)。
  • 3名ARPKD患者连续日重复肾脏MRF扫描重现性良好(T1和T2差异<3%)。

诊断标准与关键数值

  • 9个体模的MRF结果与基于自旋回波的T1和T2值高度一致(R2>0.99)。
  • 有无B1校正的肾脏T1和T2值一致性R2>0.96(P<.001)。
  • 11秒与15秒采集重建一致性R2>0.87(P<.001)。
  • ARPKD患者重复扫描T1和T2差异<3%。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background MR fingerprinting (MRF) provides rapid and simultaneous quantification of multiple tissue parameters in a single scan. Purpose To evaluate a rapid kidney MRF technique at 3.0 T in phantoms, healthy volunteers, and patients. Materials and Methods A 15-second kidney MRF acquisition was designed with 12 acquisition segments, a range of low flip angles (5°-12°), multiple magnetization preparation schema (T1, T2, and fat suppression), and an undersampled spiral trajectory. This technique was first validated in vitro using standardized T1 and T2 phantoms. Kidney T1 and T2 maps were then obtained for 10 healthy adult volunteers (mean age ± standard deviation, 35 years ± 13; six men) and three pediatric patients with autosomal recessive polycystic kidney disease (ARPKD) (mean age, 10 years ± 3; two boys) between August 2019 and October 2020 to evaluate the method in vivo. Results Results in nine phantoms showed good agreement with spin-echo-based T1 and T2 values (R2 > 0.99). In vivo MRF kidney T1 and T2 assessments in healthy adult volunteers (cortex: T1, 1362 msec ± 5; T2, 64 msec ± 5; medulla: T1, 1827 msec ± 94; T2, 69 msec ± 3) were consistent with values in the literature but with improved precision in comparison with prior MRF implementations. In vivo MRF-based kidney T1 and T2 values with and without B1 correction were in good agreement (R2 > 0.96, P < .001), demonstrating limited sensitivity to B1 field inhomogeneities. Additional MRF reconstructions using the first nine segments of the MRF profiles (11-second acquisition time) were in good agreement with the reconstructions using 12 segments (15-second acquisition time) (R2 > 0.87, P < .001). Repeat kidney MRF scans for the three patients with ARPKD on successive days also demonstrated good reproducibility (T1 and T2: <3% difference). Conclusion A kidney MR fingerprinting method provided in vivo kidney T1 and T2 maps at 3.0 T in a single breath hold with improved precision and no need for B1 correction. © RSNA, 2021 Online supplemental material is available for this article. See also the editorial by Laustsen in this issue.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. MacAskill CJ, Markley M, Farr S, et al. Rapid B1-Insensitive MR Fingerprinting for Quantitative Kidney Imaging. Radiology. 2021. PMID: 34100680. doi:10.1148/radiol.2021202302