三维打印半拟人CT体模中定量影像特征的扫描仪内可重复性
研究以48枚前列腺癌患者经PET/CT验证的淋巴结(24枚转移、24枚非转移)为模板制作定制三维打印拟人体模,用常规腹部CT协议扫描5次并以迭代重建和两种软组织核及一种骨核重建,提取影像组学特征评估可重复性;86个特征中仅19个(22%)高度可重复(CCC≥0.85),多数特征CCC<0.75,边缘增强核扫描间方差更高,PSMA阳性淋巴结特征总体更稳定。
要点
- 86个影像组学特征中仅19个(22%)扫描仪内高度可重复(CCC≥0.85)。
- 多数特征即使采集条件完全一致仍呈关键不可重复性(CCC<0.75)。
- 边缘增强核导致扫描间方差更高,PSMA阳性淋巴结特征总体更稳定。
核心结论
- 该体模研究显示,与既往体模及临床研究不同,多数影像组学特征不具备足够的扫描仪内可重复性,不能作为可靠诊断工具。
研究设计与样本
- 体模研究:以48枚前列腺癌患者经PET/CT验证的淋巴结(24枚转移、24枚非转移)为模板制作定制三维打印拟人体模,用常规腹部CT协议扫描5次,迭代重建及两种软组织核和一种骨核重建,提取影像组学特征并以一致性相关系数评估。
关键影像表现/征象
- 边缘增强核导致扫描间方差更高。
- PSMA阳性与阴性淋巴结之间检出可重复性和再现性差异,肿瘤阳性淋巴结总体特征更稳定。
诊断标准与关键数值
- 86个特征中19个(22%)高度可重复(CCC≥0.85),对重建方案易感性低。
- 多数特征CCC<0.75,即使采集条件完全一致仍呈关键不可重复性。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Radiomics has shown to provide novel diagnostic and predictive disease information based on quantitative image features in study settings. However, limited data yielded contradictory results and important questions regarding the validity of the methods remain to be answered. The purpose of this study was to evaluate how clinical imaging techniques affect the stability of radiomics features by using 3D printed anthropomorphic CT phantom to test for repeatability and reproducibility of quantitative parameters.
METHODS: 48 PET/CT validated lymph nodes of prostate cancer patients (24 metastatic, 24 non-metastatic) were used as a template to create a customized 3D printed anthropomorphic phantom. We subsequently scanned the phantom five times with a routine abdominal CT protocol. Images were reconstructed using iterative reconstruction and two soft tissue kernels and one bone kernel. Radiomics features were extracted and assessed for repeatability and susceptibility towards image reconstruction settings using concordance correlation coefficients.
RESULTS: Our analysis revealed 19 of 86 features (22 %) as highly repeatable (CCC ≥ 0.85) with low susceptibility towards image reconstruction protocols. Most features analyzed depicted critical non-repeatability with CCC's < 0.75 even under entirely consistent imaging acquisition settings. Edge enhancing kernels result in higher variances between the scans and differences in repeatability and reproducibility were detected between PSMA-positive and negative lymph nodes with overall more stable features seen in tumor positive lymph nodes.
CONCLUSIONS: Both, repeatability and reproducibility play a crucial role in the validation process of radiomics features in clinical routine. This phantom study shows that most radiomics features in contrast to previous studies, including phantom and clinical, do not depict sufficient intra-scanner repeatability to serve as reliable diagnostic tools.