文献速递腹部技术与安全CT

影像纹理、低对比度肝脏病灶可检测性及剂量影响:深度学习算法与部分模型迭代重建的比较

Image texture, low contrast liver lesion detectability and impact on dose: Deep learning algorithm compared to partial model-based iterative reconstruction.
PubMed 文献 · Eur J Radiol · 2021-06-03

在模拟非超重和超重患者的人体模体上,含6 mm直径、20HU对比度病灶,于2、6、10、15、20 mGy五个剂量水平扫描,比较深度学习重建(True Fidelity,TF)与部分模型迭代重建(ASiR-V)对噪声纹理、低对比度病灶可检测性及潜在剂量降低的影响。结果显示TF较ASiR-V更好保留类FBP噪声纹理,AUC始终更高,平均潜在剂量降低在低、中、高强度分别为7%、25%、33%。

要点

  • TF较ASiR-V更好保留类FBP噪声纹理,图像更锐利。
  • TF的AUC始终高于ASiR-V,低对比度病灶可检测性更优。
  • 高强度TF较低、中强度TF带来更大潜在剂量降低。

核心结论

  • 与ASiR-V相比,高强度TF不影响噪声纹理,并在显著更低剂量下维持低对比度肝脏病灶可检测性。

研究设计与样本

  • 人体模体研究,模拟非超重和超重患者,含6 mm直径、20HU对比度病灶,在CT系统上以2、6、10、15、20 mGy五个剂量水平扫描,采用肝脏病灶检测的临床常规方案。

关键影像表现/征象

  • TF的类FBP噪声纹理保留优于ASiR-V,表现为更小的峰值频率差和均方根偏差,图像更锐利。
  • TF的AUC始终高于ASiR-V,低对比度病灶可检测性更优。

诊断标准与关键数值

  • 噪声纹理以经眼滤镜滤波的归一化噪声功率谱表征,与FBP纹理的相似性用峰值频率差和均方根偏差评估。
  • 低对比度可检测性用通道化Hotelling观察者评估,以ROC曲线下面积(AUC)作为品质因子。
  • TF较ASiR-V的平均潜在辐射剂量降低:低强度7%、中强度25%、高强度33%。

临床应用与阅片要点

  • 在高强度TF重建下,可在显著更低剂量维持低对比度肝脏病灶可检测性。
  • TF较ASiR-V更好保留类FBP噪声纹理,图像更锐利。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare deep learning (True Fidelity, TF) and partial model based Iterative Reconstruction (ASiR-V) algorithm for image texture, low contrast lesion detectability and potential dose reduction.

METHODS: Anthropomorphic phantoms (mimicking non-overweight and overweight patient), containing lesions of 6 mm in diameter with 20HU contrast, were scanned at five different dose levels (2,6,10,15,20 mGy) on a CT system, using clinical routine protocols for liver lesion detection. Images were reconstructed using ASiR-V 0% (surrogate for FBP), 60 % and TF at low, medium and high strength. Noise texture was characterized by computing a normalized Noise Power Spectrum filtered by an eye filter. The similarity against FBP texture was evaluated using peak frequency difference (PFD) and root mean square deviation (RMSD). Low contrast detectability was assessed using a channelized Hotelling observer and the area under the ROC curve (AUC) was used as figure of merit. Potential dose reduction was calculated to obtain the same AUC for TF and ASiR-V.

RESULTS: FBP-like noise texture was more preserved with TF (PFD from -0.043mm-1 to -0.09mm-1, RMSD from 0.12mm-1 to 0.21mm-1) than with ASiR-V (PFD equal to 0.12 mm-1, RMSD equal to 0.53mm-1), resulting in a sharper image. AUC was always higher with TF than ASIR-V. In average, TF compared to ASiR-V, enabled a radiation dose reduction potential of 7%, 25 % and 33 % for low, medium and high strength respectively.

CONCLUSION: Compared to ASIR-V, TF at high strength does not impact noise texture and maintains low contrast liver lesions detectability at significant lower dose.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Racine D, Brat HG, Dufour B, et al. Image texture, low contrast liver lesion detectability and impact on dose: Deep learning algorithm compared to partial model-based iterative reconstruction. Eur J Radiol. 2021. PMID: 34120010. doi:10.1016/j.ejrad.2021.109808

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。