离体猪肝中应变弹性成像与剪切波弹性成像评估微波消融疗效

Evaluation of Microwave Ablation Efficacy by Strain Elastography and Shear Wave Elastography in ex Vivo Porcine Liver.
PubMed 文献 · Ultrasound Med Biol · 2021-06-16

研究采用离体猪肝模型,将30个样本按消融功率和时长分为3组,用超声、应变弹性成像和剪切波弹性成像评估微波消融范围。超声测量值均显著小于实际消融区,应变弹性成像除第1组SEL2外与实测无显著差异;剪切波弹性成像在中央区和边缘区消融后均显著高于消融前。结果提示应变弹性成像和剪切波弹性成像可用于评估肝组织消融疗效。

要点

  • 超声测量消融区各径线均显著小于实际消融范围。
  • 应变弹性成像测量值与实际消融区大小多无显著差异。
  • 剪切波弹性成像在消融中央区和边缘区消融后显著升高。

核心结论

  • 应变弹性成像和剪切波弹性成像可用于评估肝组织微波消融疗效。

研究设计与样本

  • 离体猪肝模型研究,30个样本按消融功率和时长分为3组:第1组(45 W,30 s)、第2组(45 W,15 s)、第3组(30 W,30 s);采用超声、应变弹性成像和剪切波弹性成像测量,以实测消融区大小为参考。

关键影像表现/征象

  • 消融后1小时,三组超声测量的横径(D1)、纵径(D2)和前后径(D3)均显著小于实测消融区(L1、L2、L3)。
  • 除第1组SEL2外,三组应变弹性成像测量值与实测消融区各径线无显著差异。
  • 剪切波弹性成像在消融中央区(a区)和边缘区(b区)消融后均显著高于消融前(所有p值<0.05)。
  • a、b、c三区消融后即刻和消融后5分钟温度均高于消融前。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The aim of this study was to evaluate the efficacy of microwave ablation by ultrasound (US), strain elastography (SE) and shear-wave elastography (SWE). An ex vivo model of porcine liver was adopted. According to ablation power and duration, 30 samples were divided into three groups: group 1 (45 W, 30 s), group 2 (45 W, 15 s) and group 3 (30 W, 30 s). US was used to measure the largest transverse diameter (D1), vertical diameter (D2) and anteroposterior diameter (D3) of the ablated area. SE was used to measure the largest transverse diameter (SEL1), vertical diameter (SEL2) and anteroposterior diameter (SEL3). The actual size of the ablated area was measured as the largest transverse diameter (L1), vertical diameter (L2) and anteroposterior diameter (L3). SWE values and temperatures were measured in the central lesion (region a), marginal area (region b) and unablated area (region c). At 1 h post-ablation, the values measured by US (D1, D2, D3) were all significantly smaller than the ablated area (L1, L2, L3) in all three groups. Except for SEL2 in group 1, there was no significant difference in the results between SEL and L among the three groups. All SWE results were significantly higher post-ablation than pre-ablation in the central lesion (region a) and marginal area (region b, all p values <0.05). In regions a, b and c, the temperatures measured immediately and 5 min post-ablation were all higher than that measured pre-ablation. These results suggest that SE and SWE can be used to evaluate the ablation efficacy of liver tissue.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Luo C, Li T, Li Z, et al. Evaluation of Microwave Ablation Efficacy by Strain Elastography and Shear Wave Elastography in ex Vivo Porcine Liver. Ultrasound Med Biol. 2021. PMID: 34140168. doi:10.1016/j.ultrasmedbio.2021.05.009

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。