低剂量腹部CT深度学习重建与基于模型的迭代重建及滤波反投影的客观与主观图像质量比较:深度学习重建更优

Superior objective and subjective image quality of deep learning reconstruction for low-dose abdominal CT imaging in comparison with model-based iterative reconstruction and filtered back projection.
PubMed 文献 · Br J Radiol · 2021-01-01

回顾性纳入90例低剂量肝脏CT患者,比较深度学习重建(DLR)、基于模型的迭代重建(MBIR)与滤波反投影(FBP)的图像质量。DLR在腹主动脉和肝实质的平均噪声低于FBP和MBIR(p < .001),CNR高于FBP和MBIR(p < .001);DLR是唯一在肥胖患者中图像噪声不高于正常BMI患者的方法。主观评价DLR噪声和整体质量优于MBIR(p < .001,p = .03)。

要点

  • 低剂量腹部CT中,DLR的腹主动脉和肝实质平均噪声低于FBP和MBIR(p < .001)。
  • DLR的腹主动脉和肝实质CNR高于FBP和MBIR(p < .001)。
  • DLR是唯一肥胖患者图像噪声不高于正常BMI患者的方法。

核心结论

  • 基于客观和主观比较,低剂量腹部CT中DLR的图像质量优于MBIR和FBP;DLR是唯一肥胖患者图像噪声不高于正常BMI患者的方法。

研究设计与样本

  • 回顾性识别连续接受低剂量肝脏CT的患者,图像分别用DLR、MBIR和FBP重建,共纳入90例(67 ± 12.7岁;63例男性;平均BMI 25.5 kg/m2),采用Dunnett检验进行统计比较。

关键影像表现/征象

  • DLR的腹主动脉和肝实质平均噪声低于FBP和MBIR(p < .001)。
  • FBP和MBIR中,肥胖患者图像噪声显著高于正常BMI患者。
  • DLR的腹主动脉和肝实质CNR高于FBP和MBIR(p < .001)。
  • 主观评价MBIR在噪声、伪影、锐利度和整体质量方面优于FBP(p < .001)。
  • 主观评价DLR在噪声(p < .001)和整体质量(p = .03)方面优于MBIR。

临床应用与阅片要点

  • 低剂量腹部CT重建时可考虑DLR以降低腹主动脉和肝实质噪声并提高CNR。
  • 肥胖患者中FBP和MBIR图像噪声较正常BMI患者增高,DLR未见此增高。
  • 主观评价DLR噪声和整体质量优于MBIR,但锐利度和伪影方面未报告与MBIR的比较结果。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: This study aimed to conduct objective and subjective comparisons of image quality among abdominal computed tomography (CT) reconstructions with deep learning reconstruction (DLR) algorithms, model-based iterative reconstruction (MBIR), and filtered back projection (FBP).

METHODS: Datasets from consecutive patients who underwent low-dose liver CT were retrospectively identified. Images were reconstructed using DLR, MBIR, and FBP. Mean image noise and contrast-to-noise ratio (CNR) were calculated, and noise, artifacts, sharpness, and overall image quality were subjectively assessed. Dunnett's test was used for statistical comparisons.

RESULTS: Ninety patients (67 ± 12.7 years; 63 males; mean body mass index [BMI], 25.5 kg/m2) were included. The mean noise in the abdominal aorta and hepatic parenchyma of DLR was lower than that in FBP and MBIR (p < .001). For FBP and MBIR, image noise was significantly higher for obese patients than for those with normal BMI. The CNR for the abdominal aorta and hepatic parenchyma was higher for DLR than for FBP and MBIR (p < .001). MBIR images were subjectively rated as superior to FBP images in terms of noise, artifacts, sharpness, and overall quality (p < .001). DLR images were rated as superior to MBIR images in terms of noise (p < .001) and overall quality (p = .03).

CONCLUSIONS: Based on objective and subjective comparisons, the image quality of DLR was found to be superior to that of MBIR and FBP on low-dose abdominal CT. DLR was the only method for which image noise was not higher for obese patients than for those with a normal BMI.

ADVANCES IN KNOWLEDGE: This study provides previously unavailable information on the properties of DLR systems and their clinical utility.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Tamura A, Mukaida E, Ota Y, et al. Superior objective and subjective image quality of deep learning reconstruction for low-dose abdominal CT imaging in comparison with model-based iterative reconstruction and filtered back projection. Br J Radiol. 2021. PMID: 34142867. doi:10.1259/bjr.20201357

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。