真实临床实践中降低剂量策略对冠状动脉CTA图像质量的影响:PROTECTION VI注册数据的亚分析
该研究基于PROTECTION VI国际多中心注册数据,纳入32个国家55个中心的3725例冠状动脉CTA患者,由核心实验室评估图像质量。结果显示,前瞻性心电触发、低管电压和迭代重建三种降低剂量策略均不是图像质量的独立预测因素,作者认为这些策略应在临床冠状动脉CTA中常规实施。
要点
- 三种降低剂量策略(前瞻性心电触发、低管电压、迭代重建)均与冠状动脉CTA图像质量无关。
- 心率(每降低10次/分,效应量0.17)和冠状动脉钙化分级(每级效应量0.15)是图像质量的独立预测因素。
- 作者认为降低剂量策略应在临床冠状动脉CTA中常规实施。
核心结论
- 这项大型国际多中心研究显示,三种降低剂量策略与冠状动脉CTA图像质量降低无关。
研究设计与样本
- PROTECTION VI注册数据的亚分析,纳入3725例患者(2109例男性、1616例女性;中位年龄59岁),来自32个国家55个中心,冠状动脉CTA图像由核心实验室审阅。
关键影像表现/征象
- 信噪比的独立预测因素包括女性(效应量2.70)、较低体重指数(每降低1单位效应量0.38)、稳定窦性心律(效应量1.71)和扫描仪制造商(效应因制造商而异)。
- 对比噪声比的独立预测因素与信噪比相同。
- 冠状动脉CTA图像质量的独立预测因素为心率(每降低10次/分效应量0.17)和冠状动脉钙化(每增加1个钙化分级效应量0.15)。
诊断标准与关键数值
- 信噪比的阅片者间一致性组内相关系数为0.874。
- 对比噪声比的阅片者间一致性组内相关系数为0.891。
- 定性评分的阅片者间一致性kappa值为0.812。
临床应用与阅片要点
- 作者认为降低剂量策略应在临床冠状动脉CTA中常规实施。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. Dose reduction strategies for coronary CTA (CCTA) have been underused in clinical practice given concern that the strategies may lower image quality. OBJECTIVE. The purpose of this study was to explore associations between dose reduction strategies and CCTA image quality in real-world clinical practice. METHODS. This subanalysis of the international Prospective Multicenter Registry on Radiation Dose Estimates of Cardiac CT Angiography in Daily Practice in 2017 (PROTECTION VI) study included 3725 patients (2109 men, 1616 women; median age, 59 years) who underwent CCTA for coronary artery evaluation performed at 55 sites in 32 countries. CCTA image sets were reviewed at a core laboratory. A range of patient and scan characteristics, including use of three dose reduction strategies (prospective ECG triggering, low tube potential, and iterative image reconstruction) and image dose, were recorded. A single core laboratory member reviewed all examinations for quantitative image quality measures, including signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), and reviewed 50% of examinations to assign a qualitative CCTA image quality score and a semiquantitative coronary calcification measure. Multivariable logistic regression models were used to identify predictors of image quality. A second core laboratory member repeated quantitative measures for 100 examinations and the qualitative measure for 383 (approximately 20%) examinations to assess interreader agreement. RESULTS. Independent predictors (p < .05) of SNR were female sex (effect size, 2.70), lower body mass index (0.38 per 1-unit decrease), stable sinus rhythm (1.71), and scanner manufacturer (variable effect across manufacturers). These factors were also the only independent predictors of CNR. Independent predictors (p < .05) of CCTA image quality were heart rate (0.17 per 10 beats/min reduction) and coronary calcification (0.15 per coronary calcification grade). None of the three dose-saving strategies or dose-length product was an independent predictor of any image quality measure. Interreader agreement analysis showed intraclass correlation coefficients of 0.874 for SNR and 0.891 for CNR and a kappa value of 0.812 for the qualitative score. CONCLUSION. This large international multicenter study shows that three dose reduction strategies were not associated with decreased CCTA image quality. CLINICAL IMPACT. The dose reduction strategies should be routinely implemented in clinical CCTA.
引用
- Stocker TJ, Nühlen N, Schmermund A, et al. Impact of Dose Reduction Strategies on Image Quality of Coronary CTA in Real-World Clinical Practice: A Subanalysis of PROTECTION VI Registry Data. AJR Am J Roentgenol. 2021. PMID: 34133193. doi:10.2214/AJR.21.26007
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。