双能CT血管成像中优化团注触发阈值联合单能图像:一项随机临床试验

Optimized Bolus Threshold for Dual-Energy CT Angiography with Monoenergetic Images: A Randomized Clinical Trial.
PubMed 文献 · Radiology · 2021-06-15

该研究先在体模中确定双能CT血管成像各keV虚拟单能图像的最佳触发阈值,再前瞻性随机比较标准碘量单能CTA、常规阈值50%减碘双能CTA和优化阈值50%减碘双能CTA;优化阈值为40 keV下30 HU,该组碘负荷最低而主动脉强化仍维持。

要点

  • 40 keV虚拟单能图像的最佳团注触发阈值为30 HU,低于常规120 kVp下100 HU参考阈值。
  • 优化阈值双能CTA组碘剂量中位数13 g,低于常规双能CTA组19 g和单能CTA组26 g。
  • 优化阈值双能CTA组主动脉CT值中位数543-610 HU,高于单能CTA组343-359 HU。

核心结论

  • 40 keV双能CT血管成像采用30 HU优化团注触发阈值,可在降低碘负荷的同时维持主动脉强化。

研究设计与样本

  • 体模研究确定各keV虚拟单能图像的最佳触发阈值;前瞻性随机临床试验连续纳入2018年8月至2019年7月行全身CTA的参与者96例(平均年龄72岁±9,80名男性),每组32例,比较单能CTA标准碘量600 mg碘/kg、常规阈值双能CTA 50%减碘300 mg碘/kg和优化阈值双能CTA 300 mg碘/kg三组。

诊断标准与关键数值

  • 以120 kVp下100 HU触发阈值为参照;体模研究确定40 keV虚拟单能图像的最佳触发阈值为30 HU。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background The bolus-tracking technique from single-energy CT has been applied to dual-energy CT (DECT) without optimization or validation. Further optimization is imperative because of a paucity of literature and differences in the attenuation profile of virtual monoenergetic images (VMIs). Purpose To determine the optimal trigger threshold with bolus-tracking technique for DECT angiography (DECTA) in a phantom study and assess the feasibility of an optimized threshold for bolus-tracking technique in DECTA at 40 keV with a 50% reduced iodine dose in human participants. Materials and Methods A phantom study with rapid kilovoltage-switching DECT was performed to determine the optimal threshold for each kiloelectron-volt VMI. In a prospective study, consecutive participants who underwent whole-body CT angiography (CTA) from August 2018 to July 2019 were randomized into three groups: single-energy CTA (SECTA) with standard iodine dose (600 mg of iodine per kilogram), DECTA with 50% reduced iodine dose (300 mg of iodine per kilogram) by using a conventional threshold, and DECTA with 300 mg of iodine per kilogram by using an optimized threshold. A trigger threshold of 100 HU at 120 kVp was used as a reference for comparison. Injected iodine doses and aortic CT numbers were compared among the three groups using Kruskal-Wallis test. Results Ninety-six participants (mean age ± standard deviation, 72 years ± 9; 80 men) were evaluated (32 participants in each group). The optimized threshold for VMIs at 40 keV was 30 HU. The median iodine dose was lower in the optimized DECTA group (13 g) compared with conventional DECTA (19 g) and SECTA (26 g) groups (P < .017 for each comparison). The median aortic CT numbers were higher in the order corresponding to conventional DECTA (655-769 HU), optimized DECTA (543-610 HU), and SECTA (343-359 HU) groups (P < .001). Conclusion The optimized trigger threshold of 30 HU for bolus-tracking technique during dual-energy CT angiography at 40 keV achieved lower iodine load while maintaining aortic enhancement. ©RSNA, 2021 Online supplemental material is available for this article. See also the editorial by Malayeri in this issue.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Noda Y, Nakamura F, Kawai N, et al. Optimized Bolus Threshold for Dual-Energy CT Angiography with Monoenergetic Images: A Randomized Clinical Trial. Radiology. 2021. PMID: 34128721. doi:10.1148/radiol.2021210102

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。