文献速递神经CT

术后非增强脑CT中外科夹与钛制骨孔盖金属伪影的适宜iMAR预设

Appropriate iMAR presets for metal artifact reduction from surgical clips and titanium burr hole covers on postoperative non-contrast brain CT.
PubMed 文献 · Eur J Radiol · 2021-06-03

纳入22例颅内动脉瘤夹闭术后行非增强脑CT的患者,比较8种iMAR预设对夹与骨孔盖伪影的定量和视觉影响。结果显示iMAR-NC对夹的伪影指数和视觉评分最低,但对骨孔盖伪影视觉评分最高;iMAR-ShI对夹和骨孔盖的视觉评分均较低,是单一预设中同时减轻两者伪影效果最好的预设。

要点

  • 对动脉瘤夹伪影,iMAR-NC预设的SD、AI和视觉评分最低(P<0.001)。
  • iMAR-NC会加重骨孔盖伪影,其骨孔盖视觉评分最高。
  • iMAR-ShI对夹和骨孔盖的视觉评分均较低,是单一预设中同时减轻两者伪影效果最好的。

核心结论

  • iMAR-NC可减轻夹的金属伪影但会加重骨孔盖伪影;术后非增强脑CT中,iMAR-ShI单一预设对夹和骨孔盖的金属伪影减轻效果最好。

研究设计与样本

  • 纳入22例颅内动脉瘤夹闭术后行非增强脑CT的患者,对图像使用8种现有iMAR预设进行后处理,以无iMAR为对照,由两名神经放射科医生对夹和骨孔盖周围进行三点视觉评分。

诊断标准与关键数值

  • 在夹周围,iMAR-NC预设的SD、AI和视觉评分最低(P<0.001);与无iMAR相比,iMAR-ShI预设的SD、AI和视觉评分显著降低(P=0.002、P=0.002、P<0.001)。在骨孔盖周围,iMAR-NC的HU、CNR和视觉评分显著高于无iMAR(P<0.001);iMAR-ShI的视觉评分在8种预设中最低(P=0.004)。视觉评分采用三点法:1=无明显伪影,2=轻度伪影,3=重度伪影。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate suitable iterative metal artifact reduction (iMAR) presets for titanium neurosurgical clips and burr hole covers (BHCs) on postoperative non-contrast computed tomography (NCCT).

METHOD: Twenty-two patients who underwent NCCT after intracranial aneurysmal clipping were included. NCCT images were postprocessed using eight currently available iMAR presets. In each image, a circular region of interest (ROI) was placed around clip, BHC, and on parietal lobe as reference. Standard deviation (SD) and attenuation value (HU) were measured in each ROI to obtain artifact index (AI) and contrast-to-noise ratio (CNR). For each iMAR preset, SD, AI, HU, and CNR were compared with those without iMAR for clips and BHCs. Visual assessment around each clip and BHC was performed by two neuroradiologists using three-point visual score (VS) (1 = no apparent, 2 = minor, and 3 = severe artifacts).

RESULTS: Among the presets, the neuro-coils preset (iMAR-NC) showed the lowest SD, AI, and VS for clips (P < 0.001). For BHCs, HU, CNR, and VS with iMAR-NC were significantly higher than those without iMAR (P < 0.001). SD, AI, and VS with the shoulder implants preset (iMAR-ShI) were significantly lower than those without iMAR for clips (P = 0.002, 0.002, and P < 0.001, respectively). For BHCs, VS with iMAR-ShI was lowest among the presets (P = 0.004).

CONCLUSIONS: Although iMAR-NC reduces metal artifacts from clips, it strengthens artifacts from BHCs. For postoperative NCCT, iMAR-ShI most effectively reduces metal artifacts from both clips and BHCs in a single preset.

来源:Eur J Radiol,查看原文。版权归原作者所有。