新生儿脑三维磁共振指纹成像:量化脑区差异与成熟度
前瞻性研究对体模和25名新生儿行三维磁共振指纹成像,空间分辨率0.7×0.7×2 mm³、采集时间不足5分钟;体模实验显示其T1、T2值与常规自旋回波序列高度相关(R²=0.998,P<0.001);新生儿枕叶白质T1、T2值低于额叶和顶叶白质(P<0.001),区域T1值与矫正胎龄呈负相关。
要点
- 三维磁共振指纹成像可在5分钟内完成新生儿全脑T1、T2定量成像,空间分辨率0.7×0.7×2 mm³。
- 新生儿枕叶白质T1、T2值低于额叶和顶叶白质,区域T1值随矫正胎龄增加而降低。
- 体模实验中三维磁共振指纹成像T1、T2值与常规自旋回波序列高度相关(R²=0.998)。
核心结论
- 快速高空间分辨率三维磁共振指纹成像可用于新生儿脑成像,其T1、T2图能够量化新生儿脑的区域差异与成熟过程。
研究设计与样本
- 前瞻性研究,纳入25名新生儿(男15名、女10名;中位矫正胎龄263天,四分位距10天),并先行体模实验验证;采用混合径向交错采集的三维磁共振指纹成像,以奇异值分解缩短重建时间、迭代重建技术改善图像质量,并与常规自旋回波序列对照。
关键影像表现/征象
- 新生儿额叶白质T1中位数2785毫秒(四分位距2684-2888毫秒)、T2中位数189.8毫秒(四分位距176.7-222.9毫秒)。
- 新生儿顶叶白质T1中位数2849毫秒(四分位距2741-2950毫秒)、T2中位数191.6毫秒(四分位距167.5-232.9毫秒)。
- 新生儿枕叶白质T1中位数2621毫秒(四分位距2513-2722毫秒)、T2中位数162.9毫秒(四分位距143.5-186.1毫秒),枕叶白质T1、T2值低于额叶和顶叶白质(P<0.001)。
诊断标准与关键数值
- 体模实验中三维磁共振指纹成像所得T1值与常规自旋回波序列高度相关(R²=0.998,P<0.001)。
- 体模实验中三维磁共振指纹成像所得T2值与常规自旋回波序列高度相关(R²=0.998,P<0.001)。
- 新生儿各脑区T1值与矫正胎龄呈负相关(系数-0.775至-0.480,P<0.05)。
临床应用与阅片要点
- 三维磁共振指纹成像采集时间不足5分钟,空间分辨率0.7×0.7×2 mm³,适用于新生儿脑T1、T2定量图。
- 阅片时注意枕叶白质T1、T2值低于额叶和顶叶白质,区域T1值随矫正胎龄增加而降低。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Magnetic resonance fingerprinting (MRF) allows the simultaneous measurement of multiple tissue properties in a single acquisition. Three-dimensional (3D) MRF with high spatial resolution can be used for neonatal brain imaging. The aim of this study is to apply 3D MRF to neonates and show regional differences and maturation in the brain.
MATERIALS AND METHODS: In this prospective study, 3D MRF using hybrid radial-interleaved acquisition was performed on phantoms and neonates from December 2019 to October 2020. For the reconstruction of 3D MRF, singular value decomposition was applied to reduce reconstruction time, and the iterative reconstruction technique was applied to improve image quality. The accuracies of T1 and T2 values derived from 3D MRF were evaluated in a phantom experiment. Regional T1 and T2 values were obtained from neonates' brain T1 and T2 maps derived from 3D MRF. Regional T1 and T2 values were compared, and their changes according to corrected gestational age were evaluated.
RESULTS: The acquisition time for 3D MRF with a spatial resolution of 0.7 × 0.7 × 2 mm3 was less than 5 minutes. The phantom study showed high correlation between T1 and T2 values derived from 3D MRF and those from conventional spin echo sequences (T1, R2 = 0.998, P < 0.001; T2, R2 = 0.998, P < 0.001). Three-dimensional MRF was performed in 25 neonates (15 boys, 10 girls; median corrected gestational age, 263 days; interquartile range, 10 days). In neonates, T1 and T2 values differed in the frontal (median [interquartile range], 2785 [2684-2888] milliseconds and 189.8 [176.7-222.9] milliseconds), parietal (2849 [2741-2950] milliseconds and 191.6 [167.5-232.9] milliseconds), and occipital white matter (2621 [2513-2722] milliseconds and 162.9 [143.5-186.1] milliseconds), showing lower values in occipital white matter (P < 0.001). Regional T1 values showed a negative relationship with corrected gestational age (coefficient, -0.775 to -0.480; P < 0.05).
CONCLUSIONS: Fast and high spatial resolution 3D MRF was applied to neonates. T1 and T2 maps derived from 3D MRF enabled the quantification of regional differences and maturation in the neonatal brain.
引用
- Yu N, Kim JY, Han D, et al. Three-Dimensional Magnetic Resonance Fingerprinting in Neonates: Quantifying Regional Difference and Maturation in the Brain. Invest Radiol. 2022. PMID: 34101674. doi:10.1097/RLI.0000000000000800
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。