仰卧位与直立位骨盆X线检查的比较:解剖、图像质量与辐射剂量的评估
60例患者在同一次检查中分别接受仰卧位和直立位骨盆X线摄影,比较解剖特征、图像质量和辐射剂量。直立位髂嵴矢状径平均增加20.6%,57例(95%)出现骨盆后倾,图像质量评分中位数从仰卧位的87%降至78%,有效剂量从仰卧位的0.12 mSv升至直立位的0.17 mSv(高47%),42例(70%)患者偏好站立位检查。作者认为仰卧位技术不能直接照搬到直立位骨盆摄影。
要点
- 直立位骨盆X线图像质量评分中位数78%,低于仰卧位的87%(p < 0.001)。
- 直立位有效剂量中位数0.17 mSv,较仰卧位0.12 mSv高47%(p < 0.001)。
- 57例(95%)患者负重位出现骨盆后倾,直立位髂嵴矢状径平均增加20.6%。
核心结论
- 负重位患者矢状径和骨盆倾斜发生改变,直立位图像整体质量下降且辐射剂量更高,仰卧位技术不能直接用于直立位骨盆摄影,需要优化采集参数。
研究设计与样本
- 60例患者在同一次检查预约中分别接受仰卧位和直立位骨盆X线摄影,测量体重指数和矢状径,采用视觉分级分析评估图像并比较骨盆倾斜,记录剂量面积乘积并输入CalDose_X软件估算有效剂量,使用MAX06和FAX06男女特异性体模进行比较。
关键影像表现/征象
- 直立位髂嵴水平矢状径中位数30.0 cm(四分位距26.0至34.0),较仰卧位24.0 cm(22.3至28.0)平均增加20.6%(p < 0.001)。
- 57例(95%)患者在负重位出现骨盆后倾。
诊断标准与关键数值
- 直立位图像质量评分中位数78%(69至85),仰卧位87%(81至91),差异p < 0.001。
- 直立位有效剂量中位数0.17 mSv(0.13至0.33),仰卧位0.12 mSv(0.08至0.18),直立位高47%(p < 0.001),受矢状径增加影响。
临床应用与阅片要点
- 42例(70%)患者偏好站立位检查。
- 作者认为直立位骨盆摄影需要优化采集参数,仰卧位技术不能直接照搬。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Pelvis radiographs are usually acquired supine despite standing imaging reflecting functional anatomy. We compared supine and erect radiographic examinations for anatomical features, radiation dose and image quality.
METHODS: 60 patients underwent pelvis radiography in both supine and erect positions at the same examination appointment. Measures of body mass index and sagittal diameter were obtained. Images were evaluated using visual grading analysis and pelvic tilt was compared. Dose-area product values were recorded and inputted into the CalDose_X software to estimate effective dose (ED). The CalDose_X software allowed comparisons using data from the erect and supine sex-specific phantoms (MAX06 & FAX06).
RESULTS: Patient sagittal diameter was greater on standing with an average 20.6% increase at the iliac crest (median 30.0, interquartile range [26.0 to 34.0] cm), in comparison to the supine position [24.0 (22.3 to 28.0) cm; p < 0.001]. 57 (95%) patients had posterior pelvic tilt on weight-bearing. Erect image quality was significantly decreased with median image quality scores of 78% (69 to 85) compared to 87% for the supine position [81 to 91] (p < 0.001). In the erect position, the ED was 47% higher [0.17 (0.13 to 0.33) mSv vs 0.12 (0.08 to 0.18) mSv (p < 0.001)], influenced by the increased sagittal diameter. 42 (70%) patients preferred the standing examination.
CONCLUSION: Patient diameter and pelvic tilt were altered on weightbearing. Erect images demonstrated an overall decrease in image quality with a higher radiation dose. Optimal acquisition parameters are required for erect pelvis radiography as the supine technique is not directly transferable.