超声探头施加皮肤压迫对颈动脉应变弹性成像的影响

Impact of Applying a Skin Compression With the Ultrasound Probe on Carotid Artery Strain Elastography.
PubMed 文献 · J Ultrasound Med · 2021-05-14

研究9名健康志愿者在颈动脉应变弹性成像中改变探头外压(预设压迫力0–5 N用于左颈总动脉,2–12 N用于左颈内动脉)对6项弹性特征及IMT、管径的影响。结果显示近壁弹性特征随预设压迫力增加而显著降低,部分远壁特征亦如此;作者建议在保证B模式图像质量前提下使用最低探头压迫以提高测量稳健性。

要点

  • 颈动脉应变弹性成像中,探头压迫力增大会使近壁弹性特征值显著降低,远壁部分特征也降低。
  • 预设压迫力变化时,72项测试中有3项近壁NIVE特征与标准实践力下的值有统计学显著差异。
  • 作者建议在保证B模式图像质量的前提下使用最低探头压迫,以提高NIVE测量的稳健性。

核心结论

  • 结果表明颈动脉应变弹性成像需要技术指南;建议在保证良好B模式图像质量时使用最低探头压迫以提高NIVE测量的稳健性。

研究设计与样本

  • 研究纳入9名健康志愿者(平均年龄43岁±13岁,6名男性),使用嵌入探头的定制声反馈手柄进行血管超声弹性成像,先记录临床标准实践力,再施加预设压迫力(左颈总动脉0–5 N,左颈内动脉2–12 N)。

关键影像表现/征象

  • 颈总动脉和颈内动脉近壁的所有弹性成像特征随预设压迫力增加而统计学显著降低,远壁半数特征也呈现该关联。
  • 在最低预设压迫力下,72项测试中有3项NIVE特征与标准实践力下的值有统计学显著差异。
  • 整体上,NIVE对探头压迫显示一定变异,线性回归斜率显示两壁在全部压迫范围内幅度变化为10.1%–45.6%。
  • 颈内动脉近壁最大IMT以及颈总动脉和颈内动脉管腔直径与预设压迫力统计学显著相关,随预设压迫力增加全程分别降低10.2%、36.2%和17.6%。
  • 其他IMT测量与施加的预设压迫力无统计学显著关联。

临床应用与阅片要点

  • 作者建议在保证良好B模式图像质量的前提下使用最低探头压迫,以提高NIVE测量的稳健性。
  • 结果表明颈动脉应变弹性成像需要技术指南。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To study the impact of varying the external compression exerted by the ultrasound probe when performing a carotid strain elastography exam.

METHODS: Nine healthy volunteers (mean age 43 years ±13 years; 6 men) underwent a vascular ultrasound elastography exam using a custom made sound feedback handle embedding the probe, and allowing the sonographer to adjust the applied compression. A clinical standard practice (SP) force was first recorded, and then predetermined compression (PDC) forces were applied, ranging from 0 to 5 N for the left common carotid artery (CCA) or 2-12 N for the left internal carotid artery (ICA). Six carotid elastography features, namely maximum and cumulated axial strains, maximum and cumulated shear strains, cumulated axial translation, and cumulated lateral translation were assessed with noninvasive vascular elastography (NIVE) on near and far walls of carotids. The carotid intima media thickness (IMT) and diameter were also measured.

RESULTS: All elastography features on the near wall of both CCA and ICA decreased statistically significantly as the PDC force increased; this association was also observed for half of the features on the far wall. Three NIVE features at the lowest PDC force (out of 72 that were tested) were statistically significantly different than values at the SP force. Overall, NIVE showed some variance to probe compression with linear regression slopes revealing changes of 10.1%-45.6% in magnitude over the whole compression range on both walls. The maximum IMT for the ICA near wall, and carotid lumen diameters of both CCA and ICA were statistically significantly associated with PDC forces; these features underwent a decrease of 10.2%, 36.2%, and 17.6%, respectively, over the whole range of PDC force increase. Other IMT measurements were not statistically significantly associated with applied PDC forces.

CONCLUSION: These results suggest the need of technical guidelines for carotid strain elastography. Using the lowest probe compression while allowing a good B-mode image quality is recommended to improve the robustness of NIVE measurements.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Chayer B, Roy Cardinal MH, Biron V, et al. Impact of Applying a Skin Compression With the Ultrasound Probe on Carotid Artery Strain Elastography. J Ultrasound Med. 2021. PMID: 33988255. doi:10.1002/jum.15750

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。