内联R2* MRI与FerriScan肝脏铁定量在祛铁治疗患者中的初步比较
内联R2*-LIC与FerriScan各勾画方式均显示良好一致性,无需离线后处理即可即时获得LIC。
要点
- 内联R2*-LIC可在扫描后即时生成肝脏铁浓度图,省去离线后处理环节。
- 与FerriScan对照,整体肝轴勾画的相关系数达94.8%、ICC 93%。
- 勾画方式会影响偏差,ROI法平均偏差约10.91%,需注意报告可比性。
核心结论
- 内联R2*-LIC与FerriScan及离线R2*校准法定量结果具有稳健且可比的一致性,无需离线后处理。
关键影像表现/征象
- 原型序列生成含与不含脂肪建模的两幅R2*图,以及基于含脂肪建模R2*图的内联R2*-LIC图。
- 勾画方式包括ROI、全肝轴层面轮廓及与FerriScan完全相同的轮廓复制。
诊断标准与关键数值
- 与FerriScan相比,全肝轴轮廓、ROI、FerriScan轮廓的内联R2*-LIC的PCC分别为94.8%、94.8%、92%,ICC分别为93%、92.7%、90.2%。
- 对应回归斜率分别为1.004、0.974、1.031,平均偏差分别为5.54%、10.91%、0.36%。
- 对应ROC-AUC分别为0.903、0.906、0.890;校正性别、年龄、诊断、肝脂肪含量及脂水交换后一致性仍维持。
临床应用与阅片要点
- 日常可考虑用内联R2*-LIC即时出具肝铁浓度,随访监测祛铁疗效。
- 报告时需注明所用勾画方式,因ROI法与FerriScan轮廓法平均偏差差异明显。
- 肝脂肪含量与脂水交换可能影响结果,判读时结合脂肪建模信息。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: MRI quantification of liver iron concentration (LIC) using R2 or R2* relaxometry requires offline post-processing causing reporting delays, administrative overhead, and added costs. A prototype 3D multi-gradient-echo pulse sequence, with inline post-processing, allows immediate calculation of LIC from an R2* map (inline R2*-LIC) without offline processing. We compared inline R2*-LIC to FerriScan and offline R2* calibration methods.
METHODS: Forty patients (25 women, 15 men; age 18-82 years), prospectively underwent FerriScan and the prototype sequence, which produces two R2* maps, with and without fat modeling, as well as an inline R2*-LIC map derived from the R2* map with fat modeling, with informed consent. For each map, the following contours were drawn: ROIs, whole-axial-liver contour, and an exact copy of contour utilized by FerriScan. LIC values from the FerriScan report and those calculated using an alternative R2 calibration were the reference standards. Results were compared using Pearson and interclass correlation coefficients (PCC, ICC), linear regression, Bland-Altman analysis, and estimation of area under the receiver operator curve (ROC-AUC).
RESULTS: Inline R2*-LIC demonstrated good agreement with the reference standards. Compared to FerriScan, inline R2*-LIC with whole-axial-liver contour, ROIs, and FerriScan contour demonstrated PCC of 94.8%, 94.8%, and 92%; ICC 93%, 92.7%, and 90.2%; regression slopes 1.004, 0.974, and 1.031; mean bias 5.54%, 10.91%, and 0.36%; and ROC-AUC estimates 0.903, 0.906, and 0.890 respectively. Agreement was maintained when adjusted for sex, age, diagnosis, liver fat content, and fat-water swap.
CONCLUSION: Inline R2*-LIC provides robust and comparable quantification of LIC compared to FerriScan, without the need for offline post-processing.
KEY POINTS: • In patients being treated for iron overload with chelation therapy, liver iron concentration (LIC) is regularly assessed in order to monitor and adjust therapy. • Magnetic resonance imaging (MRI) is commonly used to quantify LIC. Several R2 and R2* methods are available, all of which require offline post-processing. • A novel R2* MRI method allows for immediate calculation of LIC and provides comparable quantification of LIC to the FerriScan and recently published alternative R2* methods.
引用
- Healy GM, Kannengiesser SAR, Espin-Garcia O, et al. Comparison of Inline R2* MRI versus FerriScan for liver iron quantification in patients on chelation therapy for iron overload: preliminary results. Eur Radiol. 2021. PMID: 34041571. doi:10.1007/s00330-021-08019-0
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。