过压对离体大块超声消融中声发射及治疗组织组织学的影响

Effect of Overpressure on Acoustic Emissions and Treated Tissue Histology in ex Vivo Bulk Ultrasound Ablation.
PubMed 文献 · Ultrasound Med Biol · 2021-05-20

本研究用离体牛肝开展两组大块超声消融实验,以0.52或1.2 MPa静水过压抑制气泡活动与组织汽化,并设常压对照组,同步记录次谐波、宽带及低频声发射,用2% TTC染色评估热损伤范围。结果显示过压组次谐波发射显著降低,且声发射与损伤尺寸的相关性在3.1 MHz组为正、4.8 MHz组为负,提示无汽化时气泡活动可增强消融,有汽化时反而缩小热损伤范围。

要点

  • 过压可显著抑制次谐波发射,提示气泡活动受抑制。
  • 3.1 MHz时声发射与损伤尺寸正相关,4.8 MHz时负相关。
  • 气泡活动对消融的作用取决于是否发生组织汽化。

关键影像表现/征象

  • 过压组次谐波发射较对照组显著降低。
  • 3.1 MHz系列中声发射与损伤尺寸呈显著正相关。
  • 4.8 MHz系列中声发射与损伤尺寸呈显著负相关。

诊断标准与关键数值

  • 3.1 MHz组:31 W/cm2(0.9 MPa振幅)连续波作用20 min,静水过压0.52或1.2 MPa对比常压。
  • 4.8 MHz组:31 W/cm2(0.9 MPa振幅)连续波作用10 min,静水过压0.52或1.2 MPa对比常压。
  • 被动空化检测器记录次谐波(1.55或2.4 MHz)、宽带(1.2-1.5 MHz)及低频(5-20 kHz)声发射。
  • 热损伤范围以2% TTC染色评估。

临床应用与阅片要点

  • 消融中监测次谐波可反映气泡活动水平,过压可抑制气泡活动。
  • 不同频率下声发射与损伤尺寸的相关性方向相反,解读声发射信号需结合频率与汽化状态。
  • 存在组织汽化时,气泡活动可能缩小热损伤范围,需注意评估消融效果。

局限性

  • 研究为离体牛肝实验,结果不能直接外推至在体临床。
  • 仅测试两种频率和两种过压水平,未覆盖其他参数组合。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Bulk ultrasound ablation is a thermal therapy approach in which tissue is heated by unfocused or weakly focused sonication (average intensities on the order of 100 W/cm2) to achieve coagulative necrosis within a few minutes exposure time. Assessing the role of bubble activity, including acoustic cavitation and tissue vaporization, in bulk ultrasound ablation may help in making bulk ultrasound ablation safer and more effective for clinical applications. Here, two series of ex vivo ablation trials were conducted to investigate the role of bubble activity and tissue vaporization in bulk ultrasound ablation. Fresh bovine liver tissue was ablated with unfocused, continuous-wave ultrasound using ultrasound image-ablate arrays sonicating at 31 W/cm2 (0.9 MPa amplitude) for either 20 min at a frequency of 3.1 MHz or 10 min at 4.8 MHz. Tissue specimens were maintained at a static overpressure of either 0.52 or 1.2 MPa to suppress bubble activity and tissue vaporization or at atmospheric pressure for control groups. A passive cavitation detector was used to record subharmonic (1.55 or 2.4 MHz), broadband (1.2-1.5 MHz) and low-frequency (5-20 kHz) acoustic emissions. Treated tissue was stained with 2% triphenyl tetrazolium chloride to evaluate thermal lesion dimensions. Subharmonic emissions were significantly reduced in overpressure groups compared with control groups. Correlations observed between acoustic emissions and lesion dimensions were significant and positive for the 3.1-MHz series, but significant and negative for the 4.8-MHz series. The results indicate that for bulk ultrasound ablation, where both acoustic cavitation and tissue vaporization are possible, bubble activity can enhance ablation in the absence of tissue vaporization, but can reduce thermal lesion dimensions in the presence of vaporization.