团注跟踪法动态增强CT中胰腺实质期的最佳扫描延迟优化
非线性回归显示胰腺CT值与SDR显著相关(p<0.001),胰腺强化峰值(123.5 HU)对应SDR为1.00,SDR在0.89~1.18之间可获得胰腺强化≥120 HU的适当图像。结论认为TimeSD应约等于TimeTRIG以获得满意的胰腺实质期图像。
要点
- 胰腺实质期扫描延迟应个体化,使TimeSD约等于TimeTRIG(SDR≈1.00)。
- SDR在0.89~1.18范围内可保证胰腺强化≥120 HU。
- 团注跟踪触发后需根据TimeTRIG调整延迟,而非固定延迟。
核心结论
- 胰腺CT值在SDR为1.00时达峰值(123.5 HU),即TimeSD应约等于TimeTRIG以获得适当的胰腺实质期图像。
研究设计与样本
- 采用团注跟踪法,触发阈值设为腹主动脉100 HU;记录TimeTRIG(注射至触发时间)和TimeSD(触发至胰腺实质期扫描开始时间),定义SDR=TimeSD/TimeTRIG。
- 采用非线性回归分析评估胰腺CT值与SDR的关系,以胰腺CT值≥120 HU为适当强化标准。
关键影像表现/征象
- 胰腺CT值与SDR显著相关(p<0.001)。
- 胰腺强化峰值CT值为123.5 HU,对应SDR为1.00。
诊断标准与关键数值
- SDR在0.89~1.18之间对应胰腺CT值≥120 HU,视为适当的胰腺实质期强化。
- 平均TimeTRIG为16.1 s,平均TimeSD为16.8 s。
临床应用与阅片要点
- 实际扫描中应计算TimeTRIG并据此设定TimeSD,使SDR落在0.89~1.18范围内,以保证胰腺实质期强化满意。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To determine the optimal scan delay corresponding to individual hemodynamic status for pancreatic parenchymal phase in dynamic contrast-enhanced CT of the abdomen.
METHODS: One hundred and fourteen patients were included in this retrospective study (69 males and 45 females; mean age, 67.9 ± 12.1 years; range, 39-87 years). These patients underwent abdominal dynamic contrast-enhanced CT between November 2019 and May 2020. We calculated and recorded the time from contrast material injection to the bolus-tracking trigger of 100 Hounsfield unit (HU) at the abdominal aorta (s) (TimeTRIG) and scan delay from the bolus-tracking trigger to the initiation of pancreatic parenchymal phase scanning (s) (TimeSD). The scan delay ratio (SDR) was defined by dividing the TimeSD by TimeTRIG. Non-linear regression analysis was conducted to assess the association between CT number of the pancreas and SDR and to reveal the optimal SDR, which was ≥120 HU in pancreatic parenchyma.
RESULTS: The non-linear regression analysis showed a significant association between CT number of the pancreas and the SDR (p < 0.001). The mean TimeTRIG and TimeSD were 16.1 s and 16.8 s, respectively. The SDR to peak enhancement of the pancreas (123.5 HU) was 1.00. An SDR between 0.89 and 1.18 shows an appropriate enhancement of the pancreas (≥120 HU).
CONCLUSION: The CT number of the pancreas peaked at an SDR of 1.00, which means TimeSD should be approximately the same as TimeTRIG to obtain appropriate pancreatic parenchymal phase images in dynamic contrast-enhanced CT with bolus-tracking method.
ADVANCES IN KNOWLEDGE: The hemodynamic state is different in each patient; therefore, scan delay from the bolus-tracking trigger should also vary based on the time from contrast material injection to the bolus-tracking trigger. This is necessary to obtain appropriate late hepatic arterial or pancreatic parenchymal phase images in dynamic contrast-enhanced CT of the abdomen.