文献速递腹部技术与安全USMR

成人肝脏多参数超声测量的可靠性研究

Reliability of Performing Multiparametric Ultrasound in Adult Livers.
PubMed 文献 · J Ultrasound Med · 2021-05-13

该前瞻性研究在21名接受MRI-PDFF的成人中测量肝脏剪切波速度、剪切波色散斜率、衰减系数、归一化局部方差及肝肾回声强度比,发现SWV、ATI、NLV的观察者内和观察者间可靠性中等至优秀,而SWD和L/K比可靠性差。建议标准化机器设置、扫描方案和操作者培训。

要点

  • SWV、ATI、NLV的观察者内和观察者间ICC均>0.75,可靠性中等至优秀。
  • SWD在所有肝脏及L/K比在正常肝脏中的95%CI下限较低,可靠性差。
  • 进行肝脏多参数超声时需标准化机器设置、扫描方案和操作者培训。

诊断标准与关键数值

  • 正常肝脏组MRI-PDFF<5%,脂肪肝组MRI-PDFF≥5%。
  • SWV、ATI、NLV的观察者内和观察者间ICC均>0.75。
  • SWD在所有肝脏中ICC的95%CI为0.38-0.90,L/K比在正常肝脏中ICC的95%CI为0.47-0.91。
  • 脂肪肝组与正常肝脏组间SWV、SWD、ATI、NLV、L/K比及MRI-PDFF差异有统计学意义(p<.05),血清生物标志物和BMI差异无统计学意义(p>.05)。

临床应用与阅片要点

  • 测量肝脏SWV、ATI和NLV时可靠性较好,可用于临床评估。
  • 测量SWD和L/K比时可靠性差,解释结果需谨慎。
  • 建议标准化机器设置、扫描方案并进行操作者培训以提高可重复性。

局限性

  • 样本量较小(21例),可能影响结果的普遍性。
  • 未报告多中心数据,为单中心前瞻性研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The aim of the study was to test inter-observer and intra-observer reliability of measuring multiparametric ultrasound in adult livers.

METHODS: We prospectively measured shear wave velocity (SWV, m/s), shear wave dispersion slope (SWD, m/s/kHz), attenuation coefficient (ATI, dB/cm/MHz), normalized local variance (NLV), and echo intensity ratio of liver to kidney (L/K ratio) in 21 adults who underwent liver magnetic resonance imaging-proton density fat fraction (MRI-PDFF). Intraclass correlation coefficient and 95% Bland-Altman limits of agreement (95% LOA) were used to analyze intra- and inter-observer reproducibility.

RESULTS: Based on liver MRI-PDFF, 21 participants (8 men and 13 women, mean age 55 years) were divided into group 1 (11 normal livers, MRI-PDFF <5%) and group 2 (10 steatotic livers, MRI-PDFF ≥5%). ICCs for intra-observer repeatability and inter-observer reproducibility in measuring multiple ultrasound parameters in both normal and steatotic livers were above 0.75. However, 95% confidence interval for measuring SWD in all livers and L/K ratio in normal livers was 0.38-0.90 and 0.47-0.91, respectively. Differences in SWV, SWD, ATI, NLV, L/K ratio, and MRI-PDFF between participants with and without hepatic steatosis were significant (p < .05), whereas serum biomarkers and body mass index were not (p > .05), based on a two-tailed t-test.

CONCLUSIONS: The results suggest that the repeatability and reproducibility for measuring liver SWV, ATI, and NLV are moderate to excellent, while those for SWD and L/K ratio are poor. Standardized machine settings, scanning protocols, and operator training are suggested in performing multiparametric ultrasound of the liver.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Gao J, Lee R, Trujillo M, et al. Reliability of Performing Multiparametric Ultrasound in Adult Livers. J Ultrasound Med. 2021. PMID: 33982805. doi:10.1002/jum.15751

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。