高分辨率三维MRI主动脉瓣面积测量新方法及其与常规电影MRI和超声心动图评估主动脉瓣狭窄严重程度的比较
本研究纳入139例主动脉瓣狭窄患者和21例无症状对照,比较单次屏气高分辨率三维SSFP序列与常规二维电影MRI测量主动脉瓣面积(AVA)的效果,并与经胸超声心动图连续性方程测值对照。结果显示3D测量法与超声测值高度一致(ICC 0.87),观察者内一致性优于2D,观察者间一致性相似,三维法观察者变异度低。
要点
- 单次屏气高分辨率3D SSFP序列可测量AVA,观察者内一致性ICC最高达0.97(观察者2)
- 3D测AVA与超声连续性方程测值一致性ICC为0.87,略高于2D的0.85
- 3D与2D及超声三法测得的平均AVA差异无统计学意义(p=0.366)
核心结论
- 高分辨率3D MR单次屏气SSFP序列测量AVA观察者变异度低,与经胸超声心动图测值高度相关(ICC 0.87)。
- 在主动脉瓣狭窄患者中,3D、2D电影MRI与超声测得的平均AVA分别为0.81±0.26 cm²、0.82±0.34 cm²和0.80±0.26 cm²,三法差异无统计学意义(p=0.366)。
- 3D测AVA与超声的一致性略优于2D(ICC 0.87 vs. 0.85)。
研究设计与样本
- 研究纳入139例连续主动脉瓣狭窄患者(平均年龄68.5±9.4岁)和21例无症状对照(52.3±14.2岁)。
- 3D planimetry采用T2-prepared 3D SSFP序列,层厚2.0 mm、10层连续切片,于收缩中期单次屏气采集。
- 2D planimetry采用2D SSFP电影序列,层厚6.0 mm,在三个主动脉瓣水平采集。
- 以经胸超声心动图连续性方程测得的有效AVA作为参考标准进行比较。
诊断标准与关键数值
- 3D planimetry组观察者1观察者内ICC为0.95(95%CI 0.94-0.97),2D planimetry为0.87(95%CI 0.82-0.91)。
- 观察者2的观察者内ICC:3D planimetry为0.97(95%CI 0.96-0.98),2D planimetry为0.98(95%CI 0.97-0.99)。
- 3D与2D的观察者间ICC分别为0.92(95%CI 0.89-0.94)和0.91(95%CI 0.88-0.93)。
- 3D planimetry测AVA与超声连续性方程测值的ICC为0.87(95%CI 0.82-0.91),2D planimetry为0.85(95%CI 0.79-0.89)。
临床应用与阅片要点
- 可采用单次屏气高分辨率3D SSFP序列测量主动脉瓣面积,替代多层2D电影序列以减少观察者内变异。
- 3D测值与超声连续性方程测值一致性良好,可作为超声的补充或替代方法评估主动脉瓣狭窄严重程度。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: We aimed to compare the aortic valve area (AVA) calculated using fast high-resolution three-dimensional (3D) magnetic resonance (MR) image acquisition with that of the conventional two-dimensional (2D) cine MR technique.
MATERIALS AND METHODS: We included 139 consecutive patients (mean age ± standard deviation [SD], 68.5 ± 9.4 years) with aortic valvular stenosis (AS) and 21 asymptomatic controls (52.3 ± 14.2 years). High-resolution T2-prepared 3D steady-state free precession (SSFP) images (2.0 mm slice thickness, 10 contiguous slices) for 3D planimetry (3DP) were acquired with a single breath hold during mid-systole. 2D SSFP cine MR images (6.0 mm slice thickness) for 2D planimetry (2DP) were also obtained at three aortic valve levels. The calculations for the effective AVA based on the MR images were compared with the transthoracic echocardiographic (TTE) measurements using the continuity equation.
RESULTS: The mean AVA ± SD derived by 3DP, 2DP, and TTE in the AS group were 0.81 ± 0.26 cm², 0.82 ± 0.34 cm², and 0.80 ± 0.26 cm², respectively (p = 0.366). The intra-observer agreement was higher for 3DP than 2DP in one observer: intraclass correlation coefficient (ICC) of 0.95 (95% confidence interval [CI], 0.94-0.97) and 0.87 (95% CI, 0.82-0.91), respectively, for observer 1 and 0.97 (95% CI, 0.96-0.98) and 0.98 (95% CI, 0.97-0.99), respectively, for observer 2. Inter-observer agreement was similar between 3DP and 2DP, with the ICC of 0.92 (95% CI, 0.89-0.94) and 0.91 (95% CI, 0.88-0.93), respectively. 3DP-derived AVA showed a slightly higher agreement with AVA measured by TTE than the 2DP-derived AVA, with the ICC of 0.87 (95% CI, 0.82-0.91) vs. 0.85 (95% CI, 0.79-0.89).
CONCLUSION: High-resolution 3D MR image acquisition, with single-breath-hold SSFP sequences, gave AVA measurement with low observer variability that correlated highly with those obtained by TTE.
引用
- Kim HJ, Choe YH, Kim SM, et al. A New Method for Aortic Valve Planimetry with High-Resolution 3-Dimensional MRI and Its Comparison with Conventional Cine MRI and Echocardiography for Assessing the Severity of Aortic Valvular Stenosis. Korean J Radiol. 2021. PMID: 33938648. doi:10.3348/kjr.2020.1218
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。