文献速递心血管外周血管泌尿生殖CTDSA

肾移植患者一站式评估中个体化方案降低辐射与对比剂剂量

Individualized protocol for radiation and contrast medium dose reduction in one-stop assessment for kidney transplantation patients.
PubMed 文献 · Eur J Radiol · 2021-05-07

前瞻性随机纳入148例肾移植前需评估冠状动脉与髂动脉的患者,比较个体化方案与传统方案在冠脉CTA联合髂动脉CTA中的图像质量、辐射与对比剂剂量。个体化组对比剂流速降低14.4%、对比剂剂量降低35.8%、辐射剂量降低34.3%,冠脉CTA主观评分更高、噪声更低、信噪比更高,髂动脉CTA两组无差异。

要点

  • 个体化组较传统组对比剂流速3.85±0.72 mL/s vs 4.5 mL/s降低14.4%。
  • 个体化组较传统组对比剂剂量38.53±7.18 mL vs 60 mL降低35.8%。
  • 个体化组较传统组辐射剂量4.30±1.73 mSv vs 6.54±1.45 mSv降低34.3%。

核心结论

  • 肾移植前冠状动脉与髂动脉一站式评估中,个体化扫描方案在维持髂动脉CTA图像质量的同时提供更好的冠状动脉图像,并显著降低辐射与对比剂剂量。

研究设计与样本

  • 前瞻性随机对照研究,148例肾移植前需评估冠状动脉与髂动脉的患者随机分为传统组(n=72)与个体化组(n=76),均使用256排CT行一站式联合扫描,采用自动管电流调制和50%预ASIR-V控制辐射剂量;先行心跳轴向扫描模式冠脉CTA(团注跟踪技术),3秒后行螺旋扫描髂动脉CTA。

关键影像表现/征象

  • 冠脉CTA三支主要血管主观图像质量评分个体化组显著高于传统组(P<0.05)。
  • 冠脉CTA噪声个体化组17.30±4.97 HU vs 传统组19.13±4.73 HU,P=0.02。
  • 冠脉CTA信噪比个体化组22.09±7.41 vs 传统组19.55±6.18,P=0.03。
  • 髂动脉CTA主观评分与客观测量两组间均无差异。

诊断标准与关键数值

  • 传统组方案:100 kVp,350 mgI/mL对比剂60 mL,流速4.5 mL/s。
  • 个体化组管电压:BMI<24用80 kVp,BMI≥24用100 kVp。
  • 个体化组对比剂剂量:BMI<18为19 mgI/kg,18≤BMI<24为21 mgI/kg,BMI≥24为22 mgI/kg。

临床应用与阅片要点

  • 肾移植前冠状动脉与髂动脉一站式评估可采用BMI依赖的个体化管电压与对比剂剂量方案。
  • 个体化方案在降低辐射与对比剂剂量的同时可提供更好的冠状动脉图像质量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To compare image quality, radiation, and contrast medium (CM) doses between individualized and conventional scan protocols in combined coronary CT angiography (CCTA) and iliac artery CTA for kidney transplantation patients.

METHODS: 148 patients needing assessment for coronary and iliac arteries before kidney transplantation were prospectively enrolled and randomly divided into the conventional and individualized groups. All patients underwent one-stop combined scans on a 256-row CT scanner with automatic tube current modulation, 50 % pre-ASIR-V to control radiation dose. CCTA was performed first using one heartbeat axial scan mode with bolus tracking technique and iliac CTA was performed 3 s after CCTA using a spiral scan. The conventional group (n = 72) used the standard protocol: 100 kVp, 60 mL of 350 mgI/mL CM at 4.5 mL/s flow rate. The individualized group (n = 76) used a body-mass-index (BMI)-dependent protocol: kVp: 80 (BMI < 24) and 100 (BMI ≥ 24) and CM: 19 mgI/kg (BMI < 18); 21 mgI/kg (18 ≤ BMI < 24); and 22 mgI/kg (BMI ≥ 24). Image quality radiation and CM doses of the two groups were compared.

RESULTS: There was no significant difference in patient demographic data. Compared with the conventional group, the individualized group reduced contrast flow rate (in mL/s) by 14.4 % (3.85 ± 0.72 vs. 4.5), contrast dose (in mL) by 35.8 % (38.53 ± 7.18 vs. 60) and radiation dose (in mSv) by 34.3 % (4.30 ± 1.73 vs. 6.54 ± 1.45). The individualized group had significantly higher subjective image quality score (P < 0.05), lower noise (17.30 ± 4.97 HU vs. 19.13 ± 4.73 HU, P = 0.02) and higher signal-to-noise ratio (22.09 ± 7.41 vs. 19.55 ± 6.18, P = 0.03) for the three main vessels in CCTA compared with the conventional group. There were no differences in both subjective scores and objective measurements in iliac artery CTA between the two groups.

CONCLUSION: The individualized scanning protocol in the one-stop assessment of coronary and iliac arteries before kidney transplantation significantly reduces both radiation and CM doses while maintaining image quality in iliac artery CTA and providing better coronary artery images in CCTA.

来源:Eur J Radiol,查看原文。版权归原作者所有。