儿童超低剂量肺部MDCT:接近0.2毫西弗
本研究比较两种多排CT(螺旋64排与轴向宽锥256排联合深度学习重建)在儿童非增强肺部CT中的辐射剂量与图像质量。结果显示256排CT在保持主观图像质量无显著差异的同时,显著降低辐射剂量(有效剂量0.23 vs 1.13 mSv,p<0.001),但定位像剂量占比升高。结论认为采用最新扫描仪技术可将儿童肺部CT有效剂量降至0.1-0.3 mSv而不影响图像质量。
要点
- 256排宽锥CT联合深度学习重建可在儿童肺部CT中实现超低剂量(有效剂量约0.23 mSv)且主观图像质量不劣于64排CT。
- 定位像剂量占总剂量比例随总剂量降低而升高(从2.64%增至6.60%),需优化定位像扫描方案。
- 儿童CT应使用最新多排CT、体型适配曝光协议和迭代/深度学习重建以降低辐射剂量。
关键影像表现/征象
- MDCT 2的SNR和CNR高于MDCT 1,但主观图像质量评分两者无显著差异。
- MDCT 2的CTDIvol为0.26 ± 0.14 mGy,有效剂量0.23 ± 0.11 mSv;MDCT 1的CTDIvol为0.96 ± 0.52 mGy,有效剂量1.13 ± 0.58 mSv,p < 0.001。
- 定位像剂量占比从MDCT 1的2.64 ± 1.42%升高至MDCT 2的6.60 ± 5.03%(p = 0.001)。
诊断标准与关键数值
- 辐射剂量指标:CTDIvol和有效剂量,MDCT 2分别为0.26 ± 0.14 mGy和0.23 ± 0.11 mSv,MDCT 1分别为0.96 ± 0.52 mGy和1.13 ± 0.58 mSv(p < 0.001)。
- 图像质量客观指标:SNR和CNR在MDCT 2更高;主观图像质量由6名儿科放射科医生评分,两种扫描仪间无显著差异。
临床应用与阅片要点
- 在儿童肺部CT中,优先选用具备宽锥轴向采集和深度学习重建的最新多排CT,并采用体型适配曝光协议,可显著降低辐射剂量而不牺牲主观图像质量。
- 需注意定位像剂量在超低剂量方案中占比增大,应优化定位像扫描参数以进一步降低总剂量。
- 尽管MDCT 2的SNR和CNR更高,但主观评分无差异,提示在超低剂量下客观噪声指标改善不一定转化为主观感受提升。
局限性
- 样本量较小(每组23例),可能影响统计效能。
- 为单中心回顾性研究,结果外推需谨慎。
- 未报告诊断准确性或临床结局,仅评估图像质量与剂量。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare objective and subjective parameters in image quality and radiation dose of two MDCTs (helical 64 detector CT vs. axial 256 detector CT) in paediatric lung CT.
METHODS: Radiation dose and image quality were compared between non-enhanced lung CT from a helical 64-slice multidetector CT (MDCT 1) and a 256-slice scanner (MDCT 2) with axial wide-cone acquisition and using deep learning image reconstruction. In 23 size-matched paediatric studies (age 2-18 years) from each scanner, the radiation exposure, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), image sharpness and delineation of small airways were assessed. Subjective image quality was rated by 6 paediatric radiologists.
RESULTS: While MDCT 2 provided higher SNR and CNR, subjective image quality was not significantly different between studies from both scanners. Radiation exposure was lower in studies from MDCT 2 (CTDIvol 0.26 ± 0.14 mGy, effective dose 0.23 ± 0.11 mSv) than from MDCT 1 (CTDIvol 0.96 ± 0.52 mGy, effective dose 1.13 ± 0.58 mSv), p < 0.001. Despite lower radiation dose for the scout images, the relative scout-scan-ratio increased from 2.64 ± 1.42 % in MDCT 1 to 6.60 ± 5.03 % in MDCT 2 (p = 0.001).
CONCLUSIONS: By using latest scanner technology effective radiation dose can be reduced to 0.1-0.3 mSv for lung CT in children without compromising image quality. Scout image dose increasingly accounts for substantial portions of the total scan dose and needs to be optimized. In children CT should be performed on state-of-the-art MDCT scanners with size-adapted exposure protocols and iterative reconstruction.
引用
- Tschauner S, Zellner M, Pistorius S, et al. Ultra-low-dose lung multidetector computed tomography in children - Approaching 0.2 millisievert. Eur J Radiol. 2021. PMID: 33932715. doi:10.1016/j.ejrad.2021.109699
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。