标准全剂量与缩减剂量方案之间CT肺密度测量的比较

Comparison of CT Lung Density Measurements between Standard Full-Dose and Reduced-Dose Protocols.
PubMed 文献 · Radiol Cardiothorac Imaging · 2021-04-22

该回顾性分析纳入COPDGene研究中1205名接受全剂量(FD)与缩减剂量(RD)CT扫描的参与者,比较Perc15的可重复性及其预测COPD诊断与急性加重频率的能力;结果显示容积校正联合噪声降低滤波(RD-MF、RD-IR)可改善FD与RD系列扫描间肺密度测量的一致性,FD与RD的Perc15对COPD诊断均有预测价值。

要点

  • 缩减剂量扫描经中值滤波或迭代重建后,Perc15可重复性系数明显低于未滤波的缩减剂量扫描。
  • 容积校正可进一步降低全剂量与缩减剂量扫描间Perc15的测量差异。
  • 容积校正后全剂量与缩减剂量Perc15对COPD诊断的AUC分别为0.724±0.045与0.739±0.045。

核心结论

  • 容积校正与噪声降低滤波联合使用可改善连续全剂量与缩减剂量CT扫描所计算肺密度测量的可重复性。

研究设计与样本

  • 前瞻性采集数据的回顾性分析,纳入COPDGene研究中1205名参与者(平均年龄65岁±9,618名男性),于2014年11月至2017年7月间接受FD与RD CT采集方案,其中640名参与者的RD扫描另行迭代重建(IR),并对RD扫描应用中值滤波(RD-MF)。

诊断标准与关键数值

  • Perc15可重复性系数(有容积校正 vs 无容积校正):RD为29.43 HU±0.62 vs 32.81 HU±1.70;RD-MF为7.42 HU±0.42 vs 19.40 HU±2.65;RD-IR为7.10 HU±0.52 vs 22.46 HU±3.91。
  • 容积校正后FD与RD的Perc15预测COPD诊断的AUC分别为0.724±0.045与0.739±0.045。
  • 容积校正后FD与RD的Perc15预测每年≥1次急性加重的AUC分别为0.593±0.068与0.589±0.066;未行容积校正时观察到相似趋势。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the reproducibility and predicted clinical outcomes of CT-based quantitative lung density measurements using standard fixed-dose (FD) and reduced-dose (RD) scans.

MATERIALS AND METHODS: In this retrospective analysis of prospectively acquired data, 1205 participants (mean age, 65 years ± 9 [standard deviation]; 618 men) enrolled in the COPDGene study who underwent FD and RD CT image acquisition protocols between November 2014 and July 2017 were included. Of these, the RD scans of 640 participants were also reconstructed using iterative reconstruction (IR). Median filtering was applied to the RD scans (RD-MF) to investigate an alternative noise reduction strategy. CT attenuation at the 15th percentile of the lung CT histogram (Perc15) was computed for all image types (FD, RD, RD-MF, and RD-IR). Reproducibility coefficients were calculated to quantify the measurement differences between FD and RD scans. The ability of Perc15 to predict chronic obstructive pulmonary disease (COPD) diagnosis and exacerbation frequency was investigated using receiver operating characteristic analysis.

RESULTS: The Perc15 reproducibility coefficients with and without volume adjustment were as follows: RD, 29.43 HU ± 0.62 versus 32.81 HU ± 1.70; RD-MF, 7.42 HU ± 0.42 versus 19.40 HU ± 2.65; and RD-IR, 7.10 HU ± 0.52 versus 22.46 HU ± 3.91. Receiver operating characteristic curve analysis indicated that Perc15 on volume-adjusted FD and RD scans were both predictive for COPD diagnosis (area under the receiver operating characteristic curve [AUC]: FD, 0.724 ± 0.045; RD, 0.739 ± 0.045) and for having one or more exacerbation per year (AUCs: FD, 0.593 ± 0.068; RD, 0.589 ± 0.066). Similar trends were observed when volume adjustment was not applied.

CONCLUSION: A combination of volume adjustment and noise reduction filtering improved the reproducibility of lung density measurements computed using serial FD and RD CT scans.Supplemental material is available for this article.© RSNA, 2021.

来源:Radiol Cardiothorac Imaging,查看原文。版权归原作者所有。