双层探测器光谱CT用于缺血性卒中血管内治疗后鉴别血液与对比剂的首经验

Initial experience with dual-layer detector spectral CT for diagnosis of blood or contrast after endovascular treatment for ischemic stroke.
PubMed 文献 · Neuroradiology · 2021-05-27

本研究纳入63例缺血性卒中血管内治疗后一周内的患者,比较单能CT与光谱探测器CT对颅内出血或对比剂染色的诊断和分级。加入光谱数据后,8例(13%)诊断改变,4例推翻出血诊断、4例新发现出血,另有3例出血分级修改。单能CT相比光谱CT诊断颅内出血的敏感度88%、特异度87%、总准确率87%,结论认为光谱CT有助于区分颅内出血与对比剂染色。

要点

  • 血管内治疗后一周内行光谱CT,13%(8/63)的扫描颅内出血诊断发生改变。
  • 光谱CT可推翻4例出血诊断、新发现4例初始未检出的出血,并修改3例出血分级。
  • 单能CT诊断颅内出血的敏感度88%、特异度87%、总准确率87%。

核心结论

  • 血管内治疗后的光谱探测器CT有助于区分颅内出血与对比剂染色。

研究设计与样本

  • 两位阅片者评估63例缺血性卒中患者血管内治疗后一周内采集的单能CT与光谱探测器CT数据;光谱探测器CT包括平扫单色CT、碘去除后的水基图像、碘图及单能图像;参照标准为光谱探测器CT;采用Pearson卡方检验比较诊断差异,计算单能CT的诊断价值,采用Cohen Kappa评估阅片者间一致性。

关键影像表现/征象

  • 加入光谱数据后,8例(13%)扫描的颅内出血诊断发生改变:4例推翻出血诊断,4例新诊断初始未检出的出血。
  • 另有3例患者的出血分级被修改。

诊断标准与关键数值

  • 单能CT相比光谱探测器CT诊断颅内出血:敏感度88%、特异度87%、阳性诊断值88%、阴性诊断值87%、总准确率87%。
  • 颅内出血存在与否的阅片者间一致性:光谱CT为0.84(95%CI 0.51-0.86),单能CT为0.84(95%CI 0.73-0.97)。
  • 颅内出血按Heidelberg和SITS-MOST分类进行分级。

临床应用与阅片要点

  • 血管内治疗后判读颅内出血或对比剂染色时,可结合光谱探测器CT的水基图像、碘图及单能图像辅助鉴别。
  • 单能CT诊断颅内出血的敏感度88%、特异度87%,提示单能CT可能漏诊或误判出血,光谱数据可改变诊断和分级。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To determine whether spectral detector CT (SDCT) with a plain non-enhanced monochromatic CT, a water-weighted image after iodine removal, an iodine map, and Mono energetic images changes the diagnosis and classification of intracranial hemorrhage based on single energy CT after endovascular treatment (EVT) for ischemic stroke.

METHODS: Two readers evaluated single energy and SD CT data collected from 63 patients within one week after EVT. They diagnosed ICH or contrast staining, and graded ICH according to the Heidelberg and Safe Implementation of Thrombolysis in Stroke-Monitoring Study (SITS-MOST) classification. Differences in diagnosis between single energy and SD CT were tested with Pearson's chi-squared test. Diagnostic values of single energy CT were calculated. Interrater agreement was based on Cohen's Kappa.

RESULTS: When spectral data were added to single energy CT, the diagnosis of ICH changed in 8 CT scans (13%): in 4, the diagnosis of ICH was rejected and in 4, initially undetected ICH was diagnosed. In an additional 3 patients, the ICH grade was modified. CT alone had 88% sensitivity, 87% specificity, 88% positive diagnostic value, 87% negative diagnostic value, and 87% overall accuracy for ICH compared to SDCT. Interreader agreement on the presence of ICH was 0.84 (95% CI 0.51-0.86) for spectral CT and 0.84 (95% CI 0.73-0.97) for single energy CT.

CONCLUSION: SD CT after endovascular treatment contributes to the distinction between intracranial hemorrhage and contrast staining.

来源:Neuroradiology,查看原文。版权归原作者所有。