3-T胎儿磁共振在临床实践中的应用:单中心经验

Utilization of 3-T fetal magnetic resonance imaging in clinical practice: a single-institution experience.
PubMed 文献 · Pediatr Radiol · 2021-05-19

该单中心回顾性研究比较了1.5 T与3 T胎儿MRI的图像质量,由四位专科放射科医生采用改良Likert量表评分。结果显示3 T在脑部图像质量上略有优势但一致性差,且伪影更多、检查时间显著更长(75.0±15.1 min vs 55.5±13.3 min,P<0.001),总体图像质量无明确改善,因此常规临床实践中1.5 T更可取。

要点

  • 3 T胎儿MRI总体图像质量未优于1.5 T,且伪影更多、检查时间更长。
  • 评估胎儿脑部时部分医生偏好3 T,但观察者间一致性仅轻微(k=0.19)。
  • 日常胎儿MRI应优先选择1.5 T,3 T可作为备选但需注意伪影和耗时。

核心结论

  • 3 T是胎儿MRI可行的替代方案,但伪影增加和检查时间延长且总体图像质量无明确改善,常规临床实践中1.5 T更可取。

研究设计与样本

  • 采用改良Likert量表对系统、序列和伪影进行评分,并计算观察者间一致性(kappa值)。

关键影像表现/征象

  • 3 T在评估胎儿脑部时被四位医生中的两位偏好,但观察者间一致性仅轻微(k=0.19,P=0.01)。
  • 在评估胎儿胸部及腹部时,四位医生中的三位在多数病例中对1.5 T与3 T无偏好。
  • 在驻波、莫尔条纹和磁敏感伪影评估中,四位医生在多数研究中均不偏好3 T。

诊断标准与关键数值

  • 采用改良Likert量表进行图像质量评分,未报告敏感度、特异度、AUC或诊断阈值。

临床应用与阅片要点

  • 日常胎儿MRI检查应优先选择1.5 T,因其伪影较少且检查时间更短。
  • 若使用3 T,需注意驻波、莫尔条纹和磁敏感伪影对图像质量的影响,并预期检查时间延长。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: As the safety and efficacy of fetal magnetic resonance imaging (MRI) at 3 tesla (T) continues to evolve, understanding its potential benefits and limitations is becoming increasingly important.

OBJECTIVE: We aim to compare the image quality of fetal MRI between 1.5 T and 3 T in routine clinical practice.

MATERIALS AND METHODS: Fetal MRIs performed at 3 T between Jan. 1, 2019, and Dec. 31, 2019, at our institution were retrospectively reviewed by four fellowship-trained subspecialty radiologists. Imaging quality by system, sequence and artifacts were compared with matched controls at 1.5 T and rated using a modified Likert scale.

RESULTS: Thirty-three fetal MRIs at 3 T were reviewed, and a control group of studies for the same clinical indication and equivalent gestational age were selected for comparison. Two of the four radiologists preferred 3-T image quality of the brain with slight agreement among the four reviewers (k=0.19, P=0.01). Three of the four radiologists had no preference for 1.5 T vs. 3 T in the majority of cases in evaluating the chest and abdomen. In the overall assessment, 3 T was preferred in less than half of cases by all four radiologists (k=0.07, P=0.26). In the evaluation of standing wave, moire fringe and magnetic susceptibility artifacts, 3 T was not preferred in the majority of studies by all four radiologists. Total exam time was significantly longer in the 3-T fetal MRIs (75.0±15.1 min) compared to the 1.5-T fetal MRIs (55.5±13.3 min, P<0.001).

CONCLUSION: While 3 T is a feasible alternative to 1.5 T for fetal MRI, the increased artifacts and longer exam times observed at 3 T without clear improvement in overall image quality make 1.5 T preferable for fetal MRI in routine clinical practice.