四维血流MRI定量先天性分流:与有创导管检查的比较
该回顾性研究纳入33例心内或心外分流患者,比较4D flow MRI与右心导管测量分流分数(Qp/Qs)的一致性。4D flow测得的Qp/Qs与导管强相关(r=0.938)且无偏倚,RV/LV每搏量比值也与Qp/Qs强相关。结论认为4D flow MRI评估分流分数与有创导管检查结果可比。
要点
- 4D flow MRI测Qp/Qs与右心导管强相关(r=0.938)且无偏倚。
- RV/LV每搏量比值与4D flow及导管Qp/Qs均强相关(r=0.852、0.842)。
- 4D flow与Fick法心输出量仅中度相关(Qp r=0.673,Qs r=0.750)。
核心结论
- 4D flow MRI测量分流分数与有创右心导管检查结果高度一致,可作为无创评估先天性分流的参考方法。
研究设计与样本
- 以右心导管分流测量为参考,主要比较4D flow与导管估算的Qp/Qs,次要比较RV/LV每搏量比值与分流分数及心输出量。
- 统计分析采用Pearson相关和Bland-Altman图。
诊断标准与关键数值
- 4D flow测Qp/Qs与导管测值强相关,r=0.938,Bland-Altman无偏倚。
- RV/LV每搏量比值与4D flow Qp/Qs强相关(r=0.852),与导管Qp/Qs强相关(r=0.842)。
- 4D flow与导管(Fick)心输出量总体中度相关:Qp r=0.673,Qs r=0.750。
临床应用与阅片要点
- 对先天性心内或心外分流患者,可用4D flow MRI无创估算Qp/Qs,减少有创导管需求。
- RV/LV每搏量比值可作为分流分数的辅助指标,与Qp/Qs强相关。
- 4D flow与Fick法心输出量仅中度相关,解读心输出量数值时需谨慎。
局限性
- 回顾性研究,样本量较小(33例),单中心经验。
- 原文未报告4D flow与导管测量间隔时间及分流类型分层结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare invasive right heart catheterization with four-dimensional (4D) flow MRI for estimating shunt fraction in patients with intracardiac and extracardiac shunts.
MATERIALS AND METHODS: In this retrospective study, patients who underwent 4D flow MRI and invasive right heart catheterization with a shunt run between August 2015 and November 2018 were included. The primary objective was comparison of estimated shunt fraction (ratio of pulmonary-to-systemic flow, Qp/Qs) at 4D flow and catheterization. Secondary objectives included comparison of the right ventricular-to-left ventricular stroke volume ratio (RVSV/LVSV) to shunt fraction (for those with applicable shunts) and comparison of cardiac output between 4D flow and catheterization. Statistical analysis included Pearson correlation and Bland-Altman plots.
RESULTS: A total of 33 patients met inclusion criteria (mean age, 49 years ± 16 [standard deviation]; 24 women). 4D flow measurements of Qp/Qs strongly correlated with those at catheterization (r = 0.938), and there was no bias. RVSV/LVSV correlated strongly with Qp/Qs from 4D flow (r = 0.852) and catheterization (r = 0.842). Measurements of left ventricle (Qs) and right ventricle (QP) cardiac output from 4D flow and catheterization (Fick) correlated moderately overall (r = 0.673 [Qp] and r = 0.750 [Qs]).
CONCLUSION: Shunt fraction measurement using 4D flow MRI compares well with that using invasive cardiac catheterization.Supplemental material is available for this article.© RSNA, 2021.
引用
- Horowitz MJ, Kupsky DF, El-Said HG, et al. 4D Flow MRI Quantification of Congenital Shunts: Comparison to Invasive Catheterization. Radiol Cardiothorac Imaging. 2021. PMID: 33969306. doi:10.1148/ryct.2021200446
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。