验证全切片成像系统用于病理诊断的指南更新
美国病理学家协会召集专家小组,系统综述2013年原指南发布后的文献,更新全切片成像(WSI)系统诊断验证指南。建议使用至少60例验证集、建立WSI与玻璃切片观察者内诊断一致性、采用2周洗脱期,总体一致性低于95%时应特别关注。
要点
- WSI验证需至少60例病例,涵盖冰冻、活检和血液病理,排除细胞学。
- 验证应评估观察者内WSI与玻璃切片诊断一致性,两种模式间设2周洗脱期。
- 所有不一致诊断需核对,总体一致性低于95%应引起实验室特别关注。
核心结论
- 系统综述 reaffirms 使用至少60例验证集、建立观察者内诊断一致性及2周洗脱期。
- 实验室应特别关注WSI与玻璃切片总体一致性低于95%的情况。
研究设计与样本
- 美国病理学家协会召集专家小组,遵循美国国家医学院标准制定可信临床实践指南。
- 系统综述纳入冰冻切片、解剖病理标本(活检、刮宫和切除)及血液病理病例,排除细胞学。
- 采用GRADE方法重新评估和更新原指南推荐。
诊断标准与关键数值
- 验证集应包含至少60例病例,涵盖冰冻、活检和血液病理,排除细胞学。
- 需建立观察者内WSI与玻璃切片诊断一致性,两种模式间设2周洗脱期。
- 所有验证中发现的不一致诊断需核对,总体一致性低于95%时应特别关注。
临床应用与阅片要点
- 实验室验证WSI系统时应使用至少60例验证集并评估观察者内一致性。
- 验证时两种模式间应设2周洗脱期以减少记忆偏倚。
- 总体WSI-玻璃切片一致性低于95%提示需审查验证过程或系统性能。
局限性
- 指南制定基于系统综述证据,原文未报告具体研究数量或样本量。
- 排除细胞学病例,推荐不适用于细胞学WSI验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
CONTEXT.—: The original guideline, "Validating Whole Slide Imaging for Diagnostic Purposes in Pathology," was published in 2013 and included 12 guideline statements. The College of American Pathologists convened an expert panel to update the guideline following standards established by the National Academies of Medicine for developing trustworthy clinical practice guidelines.
OBJECTIVE.—: To assess evidence published since the release of the original guideline and provide updated recommendations for validating whole slide imaging (WSI) systems used for diagnostic purposes.
DESIGN.—: An expert panel performed a systematic review of the literature. Frozen sections, anatomic pathology specimens (biopsies, curettings, and resections), and hematopathology cases were included. Cytology cases were excluded. Using the Grading of Recommendations Assessment, Development, and Evaluation approach, the panel reassessed and updated the original guideline recommendations.
RESULTS.—: Three strong recommendations and 9 good practice statements are offered to assist laboratories with validating WSI digital pathology systems.
CONCLUSIONS.—: Systematic review of literature following release of the 2013 guideline reaffirms the use of a validation set of at least 60 cases, establishing intraobserver diagnostic concordance between WSI and glass slides and the use of a 2-week washout period between modalities. Although all discordances between WSI and glass slide diagnoses discovered during validation need to be reconciled, laboratories should be particularly concerned if their overall WSI-glass slide concordance is less than 95%.
引用
- Evans AJ, Brown RW, Bui MM, et al. Validating Whole Slide Imaging Systems for Diagnostic Purposes in Pathology. Arch Pathol Lab Med. 2022. PMID: 34003251. doi:10.5858/arpa.2020-0723-CP
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。