放疗兆伏级束对超声造影剂的影响

Effect of a Radiotherapeutic Megavoltage Beam on Ultrasound Contrast Agents.
PubMed 文献 · Ultrasound Med Biol · 2021-03-31

本研究比较SonoVue/Lumason与Definity/Luminity在临床兆伏级光子束(24 Gy)照射下的粒径分布与声学响应。SonoVue小于7 µm的微泡浓度下降,声衰减增加、散射减弱;Definity无明显响应。SonoVue效应在二次谐波频率最明显,脉冲反转与幅度调制可捕获(最大剂量下3.5-dB下降),有望用于受控环境下的剂量监测。

要点

  • SonoVue受兆伏级光子照射后小微泡减少、声衰减增加、散射减弱
  • Definity对放疗束无明显声学响应,效应取决于造影剂配方
  • 脉冲反转与幅度调制可捕获SonoVue二次谐波变化,或可用于剂量监测

核心结论

  • 兆伏级光子束对超声造影剂的影响取决于配方,SonoVue出现剂量相关的声学变化而Definity无显著响应。
  • SonoVue的辐射效应在二次谐波频率最明显,可用脉冲反转与幅度调制捕获(最大剂量下3.5-dB下降)。

研究设计与样本

  • 体外实验研究,比较SonoVue/Lumason与Definity/Luminity两种商用超声造影剂在临床兆伏级光子束(24 Gy)照射下的响应。
  • 以粒径分布和声学响应(衰减、散射)为观测指标,未报告样本量、中心数及具体设备/序列。

关键影像表现/征象

  • SonoVue样本中直径小于7 µm的微泡浓度下降。
  • SonoVue样本声衰减增加、声散射减弱。
  • Definity样本对辐射未表现出显著响应。

诊断标准与关键数值

  • SonoVue的辐射效应在二次谐波频率最明显,脉冲反转与幅度调制可捕获最大剂量下3.5-dB的下降。

临床应用与阅片要点

  • 使用SonoVue进行超声检查时,若患者同期接受放疗,需注意微泡声学特性可能改变。
  • 二次谐波成像结合脉冲反转与幅度调制或可用于监测SonoVue的辐射暴露,但需在良好控制环境中进一步验证。

局限性

  • 研究为体外实验,未报告样本量、重复次数及统计不确定性。
  • 仅测试两种造影剂配方和单一剂量(24 Gy),结论外推需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Collateral damage to healthy surrounding tissue during conventional radiotherapy increases when deviations from the treatment plan occur. Ultrasound contrast agents (UCAs) are a possible candidate for radiation dose monitoring. This study investigated the size distribution and acoustic response of two commercial formulations, SonoVue/Lumason and Definity/Luminity, as a function of dose on clinical megavoltage photon beam exposure (24 Gy). SonoVue samples exhibited a decrease in concentration of bubbles smaller than 7 µm, together with an increase in acoustic attenuation and a decrease in acoustic scattering. Definity samples did not exhibit a significant response to radiation, suggesting that the effect of megavoltage photons depends on the UCA formulation. For SonoVue, the influence of the megavoltage photon beam was especially apparent at the second harmonic frequency, and can be captured using pulse inversion and amplitude modulation (3.5-dB decrease for the maximum dose), which could eventually be used for dosimetry in a well-controlled environment.