双能CT:管电流设置与尿酸痛风石检测
在模拟猪生物体模中,使用单源序贯双能CT扫描,评估不同管电流组合对尿酸检出指数的影响;共分析100种管电流组合,其中5种达到检出目标,最低剂量方案为19.25/16.5 mAs(135/80 kVp),DLP 53.9 mGy*cm,检出指数2.07 mL,最高检出指数为2.11 mL,对应剂量65.3 mGy*cm、8.25/79.75 mAs;最优管电流比(TCR)为2至4,过高或过低都会降低检出指数。
要点
- 最优TCR为2至4,过高或过低均降低尿酸检出指数。
- 高能扫描需达到最低管电流,才能获得可接受的总体敏感度。
- TCR高于10无益,设计DECT协议时需特别注意管电流配置。
核心结论
- 高能扫描需达到最低管电流才能获得可接受的总体敏感度,之后增加低能曝光才会检出更多尿酸;TCR高于10无益,最优TCR为2至4,说明设计合适的DECT协议须特别谨慎。
研究设计与样本
- 在模拟猪生物体模中,对含不同尿酸浓度的标本进行单源序贯双能CT重复扫描,80 kVp(16.5-220 mA s)与135 kVp(2.75-19.25 mA s)配对,分析全部100种管电流组合的检出指数。
诊断标准与关键数值
- 检出指数定义为真阳性尿酸体积减去假阳性尿酸体积;检出目标为达到全部组合最高检出指数的至少85%,且等效单能辐射剂量不超过150%。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To derive optimal scanning parameters for single-source dual-energy computed tomography (DECT) in the detection of urate by analyzing influence of tube current ratio (TCR) and total radiation exposure in a phantom.
METHOD: Specimens with different urate concentrations in a realistic porcine bio-phantom were repeatedly imaged with sequential single-source DECT scans at 80 kVp (16.5-220 mA s) and 135 kVp (2.75-19.25 mA s). Detection index (DI - true positive minus false positive urate volume) was calculated for every possible tube current combination. Optimal tube current combinations reaching at least 85 % of the highest measured DI of all combinations without exceeding 150 % of equivalent single-energy radiation dose were identified. TCR, DLP and DI were plotted and compared.
RESULTS: Cubic regression analysis showed a flattening increase in the DI with increasing tube currents. Five out of the 100 tube current combinations analyzed achieved the detection target: the lowest DLP of 53.9 mGy*cm at 19.25/16.5 mAs (135/80 kVp) achieved a DI of 2.07 mL and the highest DI of 2.11 mL at a dose of 65.3 mGy*cm and 8.25/79.75 mAs. The optimal TCR is between two and four, while both, higher and lower ratios decreased DI.
CONCLUSIONS: A minimum tube current of the high-energy scans is needed before an acceptable overall sensitivity is achieved and before increases in low-energy exposure result in more urate detection. High TCRs above 10 are not beneficial while the optimal TCR ranges between two and four, indicating that special care has to be taken in designing a suitable DECT protocol.
引用
- Kotlyarov M, Hermann KGA, Mews J, et al. Dual-energy computed tomography: Tube current settings and detection of uric acid tophi. Eur J Radiol. 2021. PMID: 33839428. doi:10.1016/j.ejrad.2021.109692