临床碲化镉光子计数CT探测器在尸体腕关节成像中的图像质量评估
该研究对比了临床碲化镉光子计数探测器CT(PCD-CT)与第三代双源能量积分探测器CT(EID-CT)在8例尸体腕关节中的成像性能,采用等效辐射剂量协议(低/标准/全剂量:CTDIvol=1.50/5.80/8.67 mGy),并由3名放射科医师对图像质量进行7分制主观评分。结果显示,标准剂量PCD-CT超高分辨率模式的图像质量优于全剂量PCD-CT标准分辨率模式(P=0.016)和全剂量EID-CT(P=0.040);低剂量PCD-CT超高分辨率模式与标准剂量PCD-CT标准分辨率模式或EID-CT无差异。PCD-CT的信噪比和对比噪声比显著高于EID-CT(所有P<0.001)。结论认为,临床PCD-CT在尸体腕关节扫描中能更优显示精细解剖,超高分辨率模式提示可显著降低剂量。
要点
- 标准剂量PCD-CT超高分辨率模式图像质量优于全剂量EID-CT和全剂量PCD-CT标准分辨率模式。
- 低剂量PCD-CT超高分辨率模式与标准剂量EID-CT或PCD-CT标准分辨率模式图像质量无差异。
- PCD-CT在等效剂量下信噪比和对比噪声比显著高于EID-CT。
局限性
- 研究基于尸体腕关节,样本量较小(8例),结论需在临床活体研究中进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Detailed visualization of bone microarchitecture is essential for assessment of wrist fractures in computed tomography (CT). This study aims to evaluate the imaging performance of a CT system with clinical cadmium telluride-based photon-counting detector (PCD-CT) compared with a third-generation dual-source CT scanner with energy-integrating detector technology (EID-CT).
MATERIALS AND METHODS: Both CT systems were used for the examination of 8 cadaveric wrists with radiation dose equivalent scan protocols (low-/standard-/full-dose imaging: CTDIvol = 1.50/5.80/8.67 mGy). All wrists were scanned with 2 different operating modes of the photon-counting CT (standard-resolution and ultra-high-resolution). After reformatting with comparable reconstruction parameters and convolution kernels, subjective evaluation of image quality was performed by 3 radiologists on a 7-point scale. For estimation of interrater reliability, we report the intraclass correlation coefficient (absolute agreement, 2-way random-effects model). Signal-to-noise and contrast-to-noise ratios were calculated to provide semiquantitative assessment of image quality.
RESULTS: Subjective image quality of standard-dose PCD-CT examinations in ultra-high-resolution mode was superior compared with full-dose PCD-CT in standard-resolution mode (P = 0.016) and full-dose EID-CT (P = 0.040). No difference was ascertained between low-dose PCD-CT in ultra-high-resolution mode and standard-dose scans with either PCD-CT in standard-resolution mode (P = 0.108) or EID-CT (P = 0.470). Observer evaluation of standard-resolution PCD-CT and EID-CT delivered similar results in full- and standard-dose scans (P = 0.248/0.509). Intraclass correlation coefficient was 0.876 (95% confidence interval, 0.744-0.925; P < 0.001), indicating good reliability. Between dose equivalent studies, signal-to-noise and contrast-to-noise ratios were substantially higher in photon-counting CT examinations (all P's < 0.001).
CONCLUSIONS: Superior visualization of fine anatomy is feasible with the clinical photon-counting CT system in cadaveric wrist scans. The ultra-high-resolution scan mode suggests potential for considerable dose reduction over energy-integrating dual-source CT.