四维多期稳态MRI联合铁氧体增强:儿童先天性心脏病的早期多中心可行性研究
本前瞻性多中心研究评估了四维MUSIC MRI在儿童先天性心脏病中的可行性。共分析60例(每中心20例),93%图像质量良好或优秀,94%心内及血管连接诊断信心高或明确,且与2D电影序列相关性高(r=0.72-0.85),同时显著缩短采集时间(12±3分钟 vs 44±20分钟)。
要点
- 四维MUSIC MRI在儿童先心病多中心应用中图像质量良好,诊断信心高。
- 该技术可动态显示全部解剖特征,简化采集流程并显著缩短检查时间。
- 与2D电影序列测量的心室容积和射血分数相关性高(r=0.72-0.85)。
关键影像表现/征象
- 56/60(93%)检查图像质量评分为良好或优秀,覆盖不同疾病复杂程度。
- 伪影评分与图像质量呈负相关(r=-0.88,P<.001),与呼吸运动抑制也呈负相关(r=-0.45,P<.001)。
- 780个心内及血管连接中730个(94%)诊断信心为高或明确。
诊断标准与关键数值
- 图像质量采用五点量表评分,基于伪影、图像质量和诊断信心。
- 四维MUSIC与2D电影序列测量的心室容积和射血分数相关性高(r=0.72-0.85,P<.001)。
- 四维MUSIC图像采集时间12分钟±3,显著短于局部标准MRI的44分钟±20(P<.001)。
临床应用与阅片要点
- 四维MUSIC可动态评估先心病心内及血管连接,减少对专家操作的依赖。
- 图像质量受伪影影响大,需注意呼吸运动抑制以优化图像。
- 与2D电影序列心室容积和射血分数相关性高,可用于心功能评估。
局限性
- 仅从460项检查中选取60例进行详细分析,样本量有限。
- 研究为前瞻性多中心可行性研究,未报告诊断准确性的金标准对比。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background The value of MRI in pediatric congenital heart disease (CHD) is well recognized; however, the requirement for expert oversight impedes its widespread use. Four-dimensional (4D) multiphase steady-state imaging with contrast enhancement (MUSIC) is a cardiovascular MRI technique that uses ferumoxytol and captures all anatomic features dynamically. Purpose To evaluate multicenter feasibility of 4D MUSIC MRI in pediatric CHD. Materials and Methods In this prospective study, participants with CHD underwent 4D MUSIC MRI at 3.0 T or 1.5 T between 2014 and 2020. From a pool of 460 total studies, an equal number of MRI studies from three sites (n = 60) was chosen for detailed analysis. With use of a five-point scale, the feasibility of 4D MUSIC was scored on the basis of artifacts, image quality, and diagnostic confidence for intracardiac and vascular connections (n = 780). Respiratory motion suppression was assessed by using the signal intensity profile. Bias between 4D MUSIC and two-dimensional (2D) cine imaging was evaluated by using Bland-Altman analysis; 4D MUSIC examination duration was compared with that of the local standard for CHD. Results A total of 206 participants with CHD underwent MRI at 3.0 T, and 254 participants underwent MRI at 1.5 T. Of the 60 MRI examinations chosen for analysis (20 per site; median participant age, 14.4 months [interquartile range, 2.3-49 months]; 33 female participants), 56 (93%) had good or excellent image quality scores across a spectrum of disease complexity (mean score ± standard deviation: 4.3 ± 0.6 for site 1, 4.9 ± 0.3 for site 2, and 4.6 ± 0.7 for site 3; P < .001). Artifact scores were inversely related to image quality (r = -0.88, P < .001) and respiratory motion suppression (P < .001, r = -0.45). Diagnostic confidence was high or definite in 730 of 780 (94%) intracardiac and vascular connections. The correlation between 4D MUSIC and 2D cine ventricular volumes and ejection fraction was high (range of r = 0.72-0.85; P < .001 for all). Compared with local standard MRI, 4D MUSIC reduced the image acquisition time (44 minutes ± 20 vs 12 minutes ± 3, respectively; P < .001). Conclusion Four-dimensional multiphase steady-state imaging with contrast enhancement MRI in pediatric congenital heart disease was feasible in a multicenter setting, shortened the examination time, and simplified the acquisition protocol, independently of disease complexity. Clinical trial registration no. NCT02752191 © RSNA, 2021 Online supplemental material is available for this article. See also the editorial by Roest and Lamb in this issue.
引用
- Nguyen KL, Ghosh RM, Griffin LM, et al. Four-dimensional Multiphase Steady-State MRI with Ferumoxytol Enhancement: Early Multicenter Feasibility in Pediatric Congenital Heart Disease. Radiology. 2021. PMID: 33876971. doi:10.1148/radiol.2021203696
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。