光子计数探测器CT定位像评估骨密度:体模尺寸和管电压的影响

Assessment of Bone Mineral Density From a Computed Tomography Topogram of Photon-Counting Detector Computed Tomography-Effect of Phantom Size and Tube Voltage.
PubMed 文献 · Invest Radiol · 2021-01-01

研究在离体骨质疏松猪腰椎模型上,评估光子计数探测器CT定位像在不同模拟尺寸和管电压下测量骨密度的准确性。结果显示,与双能X线吸收法相比,光子计数探测器CT定位像在5种尺寸和2种管电压下测量的骨密度均无显著差异(P=0.7),且骨密度从L2到L4显著降低(P<0.0001)。

要点

  • 光子计数探测器CT定位像可定量骨密度,与双能X线吸收法准确性相似。
  • 模拟尺寸和管电压对骨密度测量无显著影响(P=0.7和P=0.67)。
  • 骨密度从L2到L4显著降低(所有P<0.0001)。

核心结论

  • 在离体骨质疏松动物脊柱中,光子计数探测器CT单次定位像(2个能量阈值)在多种模拟尺寸和管电压下,骨密度定量准确性相似于双能X线吸收法。

研究设计与样本

  • 离体动物实验:猪腰椎模拟骨质疏松,5种脂肪层厚度(每层3 cm)模拟5种尺寸,分别用双能X线吸收法(参考标准)、光子计数探测器CT 120 kVp(阈值20和70 keV)和140 kVp(阈值20和75 keV)成像,测量L2-L4骨密度,线性混合效应模型分析。

诊断标准与关键数值

  • 骨密度测量值:双能X线吸收法0.56±0.03 g/cm²,光子计数探测器CT 120 kVp 0.55±0.02 g/cm²,140 kVp 0.55±0.02 g/cm²(均值±标准差,平均3个椎体和5种尺寸)。L2-L4骨密度:双能X线吸收法0.59±0.01、0.57±0.01、0.52±0.02 g/cm²;120 kVp 0.57±0.02、0.55±0.01、0.53±0.01 g/cm²;140 kVp 0.57±0.01、0.55±0.01、0.53±0.01 g/cm²。成像设置影响P=0.7,模拟尺寸影响P=0.67,交互项均P>0.2。

临床应用与阅片要点

  • 光子计数探测器CT定位像或可用于骨密度评估,但作者结论为在离体模型中与双能X线吸收法准确性相似。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The aim of this study was to assess the accuracy and impact of different sizes and tube voltages on bone mineral density (BMD) assessment using a computed tomography (CT) topogram acquired with photon-counting detector CT in an osteopenic ex vivo animal spine.

MATERIALS AND METHODS: The lumbar back of a piglet was used to simulate osteopenia of the lumbar spine. Five fat layers (each with a thickness of 3 cm) were consecutively placed on top of the excised spine to emulate a total of 5 different sizes. Each size was repeatedly imaged on (A) a conventional dual-energy x-ray absorptiometry scanner as the reference standard, (B) a prototype photon-counting detector CT system at 120 kVp with energy thresholds at 20 and 70 keV, and (C) the same prototype system at 140 kVp with thresholds at 20 and 75 keV. Material-specific data were reconstructed from spectral topograms for B and C. Bone mineral density was measured for 3 lumbar vertebrae (L2-L4). A linear mixed-effects model was used to estimate the impact of vertebra, imaging setup, size, and their interaction term on BMD.

RESULTS: The BMD of the lumbar spine corresponded to a T score in humans between -4.2 and -4.8, which is seen in osteoporosis. Averaged across the 3 vertebrae and 5 sizes, mean BMD was 0.56 ± 0.03, 0.55 ± 0.02, and 0.55 ± 0.02 g/cm2 for setup A, B, and C, respectively. There was no significant influence of imaging setup (P = 0.7), simulated size (P = 0.67), and their interaction term (both P > 0.2) on BMD. Bone mineral density decreased significantly from L2 to L4 for all 3 setups (all P < 0.0001). Bone mineral density was 0.59 ± 0.01, 0.57 ± 0.01, and 0.52 ± 0.02 g/cm2 for L2, L3, and L4, respectively, for setup A; 0.57 ± 0.02, 0.55 ± 0.01, and 0.53 ± 0.01 g/cm2 for setup B; and 0.57 ± 0.01, 0.55 ± 0.01, and 0.53 ± 0.01 g/cm2 for setup C.

CONCLUSION: A single CT topogram acquired on photon-counting detector CT with 2 energy thresholds enabled BMD quantification with similar accuracy compared with dual-energy x-ray absorptiometry over a range of simulated sizes and tube voltages in an osteopenic ex vivo animal spine.