全自动三维MRI分析膝关节软骨厚度与投影软骨面积比的扫描间测量误差

Interscan measurement error of knee cartilage thickness and projected cartilage area ratio at 9 regions and 45 subregions by fully automatic three-dimensional MRI analysis.
PubMed 文献 · Eur J Radiol · 2021-04-08

本研究使用基于深度神经网络的全自动三维MRI分析软件,对10名健康志愿者同日两次扫描的膝关节软骨厚度和投影软骨面积比(厚度≥1.5 mm)在9个区域和45个亚区的扫描间测量误差进行评估。结果显示软骨厚度测量误差在所有9个区域和45个亚区中的39个小于0.10 mm,误差范围0.03–0.21 mm;投影软骨面积比误差在9个区域为0.01–0.03。

要点

  • 全自动软件在9个区域和45个亚区中的39个软骨厚度扫描间误差<0.10 mm。
  • 软骨厚度ICC在9个区域和41个亚区均>0.75,重复性良好。
  • 投影软骨面积比扫描间误差在9个区域为0.01–0.03。

核心结论

  • 全自动三维MRI分析软件测量膝关节软骨厚度和投影软骨面积比的扫描间误差较小,可用于纵向评估。

研究设计与样本

  • 研究纳入10名健康志愿者,同日进行两次MRI扫描,采用基于深度神经网络的全自动三维MRI分析软件分割膝关节软骨。
  • 扫描间测量误差基于9名无体动志愿者的数据计算,排除1名扫描期间有体动的志愿者。
  • 软件自动输出9个区域和45个亚区的软骨厚度及投影软骨面积比(厚度≥1.5 mm),无需手动校正。

诊断标准与关键数值

  • 软骨厚度扫描间测量误差在所有9个区域和45个亚区中的39个小于0.10 mm,误差范围0.03–0.21 mm。
  • 软骨厚度ICC在所有9个区域和41个亚区高于0.75。
  • 投影软骨面积比扫描间测量误差在所有9个区域为0.01–0.03。

临床应用与阅片要点

  • 全自动软件无需手动校正即可输出多区域软骨厚度和投影软骨面积比,适合纵向随访。
  • 扫描间误差<0.10 mm提示厚度变化超过该范围可能具有临床意义,但需结合具体区域判断。
  • 投影软骨面积比误差较小(0.01–0.03),可作为软骨形态变化的补充指标。

局限性

  • 样本量小(10名健康志愿者,分析9名),仅评估健康人群。
  • 仅评估同日两次扫描的扫描间误差,未涉及不同设备或序列。
  • 排除1名有体动的志愿者,可能低估实际临床扫描中的误差。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: We have developed a fully automatic three-dimensional MRI analysis software program for automatic segmentation of knee cartilage using a deep neural network. The purpose of this study was to use this software to clarify the interscan measurement error of the knee cartilage thickness and projected cartilage area ratio at 9 regions and 45 subregions in the knee.

METHODS: Ten healthy volunteers underwent MRI twice in the same day. The software provided cartilage thickness and projected cartilage area ratio (thickness ≥ 1.5 mm) at 9 regions and 45 subregions of the knee without any manual correction. The interscan measurement error was calculated at each region and subregion from the data of nine donors, except for one donor who had body motion during the MRI examination.

RESULTS: The interscan measurement error of cartilage thickness was less than 0.10 mm at all 9 regions and at 39 subregions among 45 subregions. The measurement errors ranged from 0.03 to 0.21 mm. The intraclass correlation coefficients (ICC) of cartilage thickness were higher than 0.75 at all 9 regions and 41 subregions. The interscan measurement error of the projected cartilage area ratio ranged from 0.01 to 0.03 for all 9 regions.

CONCLUSIONS: This study clarified the interscan measurement error of the knee cartilage thickness and projected cartilage area ratio.

来源:Eur J Radiol,查看原文。版权归原作者所有。