控制层间化学位移伪影以改善7T非脂肪抑制肌肉骨骼涡轮自旋回波MRI
本研究在7T系统上探讨二维涡轮自旋回波(TSE)成像中层间化学位移伪影的控制,通过匹配并提高激发与重聚焦射频脉冲带宽(MIB序列,1500 Hz)并与传统不匹配中等带宽(UMB序列,1095和682 Hz)对比,在模型和健康志愿者膝关节成像中验证。结果表明MIB序列显著减小骨-软骨界面化学位移伪影宽度,减轻伪影对图像质量的负面影响,且信噪比与UMB相当。
要点
- 7T非脂肪抑制TSE中层间化学位移伪影可类似骨折或骨 marrow 病变,需警惕。
- 采用匹配的高射频脉冲带宽(1500 Hz)可显著减少骨-软骨界面化学位移伪影。
- MIB序列图像质量优于UMB,脂肪信号明亮,精细结构显示良好,信噪比相当。
核心结论
- 在7T系统上,对于非脂肪抑制的二维TSE成像,选择高且匹配的激发和重聚焦射频脉冲带宽(MIB)比传统不匹配较低带宽(UMB)序列 consistently 获得更好的图像质量。
研究设计与样本
- 前瞻性研究,经当地伦理委员会批准,在油-水模型及10名健康志愿者中比较MIB与UMB序列的T1加权矢状位膝关节图像,由2名放射科医生独立评估。
关键影像表现/征象
- UMB序列图像在所有检查区域出现大量伪影性高信号和弥漫性及局部严重的脂肪信号丢失,而MIB序列 consistently 产生高图像质量,T1加权脂肪信号明亮,精细组织结构显示极佳。
诊断标准与关键数值
- 骨-软骨界面化学位移伪影宽度:位置1 MIB 0.35 ± 0.20 mm vs UMB 1.27 ± 0.27 mm(P < 0.001);位置2 MIB 0.25 ± 0.13 mm vs UMB 1.48 ± 0.46 mm(P < 0.001);组内相关系数=0.98。
- 化学位移伪影对图像质量的负面影响评分(5分制):骨 MIB 2 ± 0 vs UMB 4 ± 1(P < 0.004,两名读者);肌肉 MIB 3 ± 0 vs UMB 2 ± 0(P < 0.004,两名读者);κ=0.69。
- UMB与MIB序列的信噪比相当,比值为99 ± 7%。
临床应用与阅片要点
- 在7T非脂肪抑制TSE肌肉骨骼成像中,应选择匹配的高射频脉冲带宽(如1500 Hz)以减少层间化学位移伪影,避免将伪影误判为骨折或骨 marrow 病变。
局限性
- 研究仅纳入10名健康志愿者,样本量小;仅评估了膝关节,并在单一受试者中获取腕和足图像,结论普适性有限。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Through-slice chemical shift artifacts in state-of-the-art turbo-spin-echo (TSE) images can be significantly more severe at 7 T than at lower field strengths. In musculoskeletal applications, these artifacts appear similar to bone fractures or neoplastic bone marrow disease. The objective of this work was to explore and reduce through-slice chemical shift artifacts in 2-dimensional (2D) TSE imaging at 7 T.
MATERIALS AND METHODS: This prospective study was approved by the local ethics board. The bandwidths of the excitation and refocusing radiofrequency (RF) pulses of a prototype 2D TSE sequence were individually modified and their effect on the slice profiles and relative slice locations of water and fat spins was assessed in an oil-water phantom. Based on these results, it was hypothesized that the combination of matched and increased excitation and refocusing RF pulse bandwidths ("MIB") of 1500 Hz would enable 2D TSE imaging with significantly reduced chemical shift artifacts compared with a state-of-the-art sequence with unmatched and moderate RF pulse bandwidths ("UMB") of 1095 and 682 Hz.A series of T1-weighted sagittal knee examinations in 10 healthy human subjects were acquired using the MIB and UMB sequences and independently evaluated by 2 radiologists. They measured the width of chemical shift artifacts at 2 standardized locations and graded the perceived negative effect of chemical shift artifacts on image quality in the bones and in the whole gastrocnemius muscle on a 5-point scale. Similar knee, wrist, and foot images were acquired in a single subject. Signal-to-noise ratios in the femoral bone marrow were computed between the UMB and MIB sequences.
RESULTS: Phantom measurements confirmed the expected spatial separation of simultaneously affected water and fat slices between 40% and 200% of the prescribed slice thickness for RF pulse bandwidths between 2500 and 500 Hz. Through-slice chemical shift artifacts at the bone-cartilage interface were significantly smaller with MIB than with UMB (location 1: 0.35 ± 0.20 mm vs 1.27 ± 0.27 mm, P < 0.001; location 2: 0.25 ± 0.13 mm vs 1.48 ± 0.46 mm, P < 0.001; intraclass correlation coefficient = 0.98). The negative effect of chemical shift artifacts on image quality was significantly smaller with MIB than with UMB (bone: 2 ± 0 vs 4 ± 1, P < 0.004 [both readers]; muscle: 3 ± 0 vs 2 ± 0, P < 0.004 [both readers]; κ = 0.69). The signal-to-noise ratio of the UMB and MIB sequences was comparable, with a ratio of 99 ± 7%. Images acquired using the UMB sequence displayed numerous artifactual hyperintensities and diffuse, as well as locally severe, fat signal loss in all examined regions, whereas the MIB sequence consistently yielded high image quality with bright T1-weighted fat signal and excellent depiction of fine tissue structures.
CONCLUSIONS: On 7 T systems, the selection of high and matched RF bandwidths for excitation and refocusing pulses for 2D TSE imaging without fat suppression showed consistently better image quality than state-of-the-art sequences with unmatched lower RF pulse bandwidths.