文献速递乳腺技术与安全乳腺X线

数字乳腺X线摄影影像质量评估工具的ROC验证

Validation of a candidate instrument to assess image quality in digital mammography using ROC analysis.
PubMed 文献 · Eur J Radiol · 2021-03-30

本研究用ROC分析验证一个候选影像质量评估工具在数字乳腺X线摄影中的有效性。结果显示该工具与ROC分析结果相关,可用于不同专业人员评估临床相关影像质量差异。

要点

  • 候选影像质量评估工具的视觉分级评分与ROC检测性能排名一致。
  • 放射科医生与物理师的评估结果相似,平均差异约10%。
  • 该工具可用于放射科医生、技师和物理师评估DM影像质量。

核心结论

  • 候选评估工具的视觉分级分析结果与ROC分析检测性能结果相关,可用于不同专业人员评估数字乳腺X线摄影的临床相关影像质量变化。

研究设计与样本

  • 视觉分级分析(VGA)研究使用候选工具评估其中25张图像的子集,由参与ROC研究的8名放射科医生和7名乳腺影像物理师完成。

诊断标准与关键数值

  • 四个影像质量等级的排序一致:等级1(部分AUCROC 0.070,VGAS 6.77)优于等级2(0.066,6.15)、等级3(0.061,5.82)和等级4(0.062,5.37)。
  • 放射科医生与物理师的评估结果相似,平均VGAS差异为10%。

临床应用与阅片要点

  • 该候选工具可用于放射科医生、技师和物理师评估DM影像质量的临床相关变化。
  • 评估时需注意不同专业人员的评分可能存在约10%的平均差异。

局限性

  • ROC和VGA研究均未检测到四个影像质量等级间的较大差异。
  • VGA研究仅评估了ROC研究中25张图像的子集。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To validate a candidate instrument, to be used by different professionals to assess image quality in digital mammography (DM), against detection performance results.

METHODS: A receiver operating characteristics (ROC) study was conducted to assess the detection performance in DM images with four different image quality levels due to different quality issues. Fourteen expert breast radiologists from five countries assessed a set of 80 DM cases, containing 60 lesions (40 cancers, 20 benign findings) and 20 normal cases. A visual grading analysis (VGA) study using a previously-described candidate instrument was conducted to evaluate a subset of 25 of the images used in the ROC study. Eight radiologists that had participated in the ROC study, and seven expert breast-imaging physicists, evaluated this subset. The VGA score (VGAS) and the ROC and visual grading characteristics (VGC) areas under the curve (AUCROC and AUCVGC) were compared.

RESULTS: No large differences in image quality among the four levels were detected by either ROC or VGA studies. However, the ranking of the four levels was consistent: level 1 (partial AUCROC: 0.070, VGAS: 6.77) performed better than levels 2 (0.066, 6.15), 3 (0.061, 5.82), and 4 (0.062, 5.37). Similarity between radiologists' and physicists' assessments was found (average VGAS difference of 10 %).

CONCLUSIONS: The results from the candidate instrument were found to correlate with those from ROC analysis, when used by either observer group. Therefore, it may be used by different professionals, such as radiologists, radiographers, and physicists, to assess clinically-relevant image quality variations in DM.

来源:Eur J Radiol,查看原文。版权归原作者所有。