乳腺X线摄影中自我压迫对压迫力、乳腺厚度和平均腺体剂量的影响
本研究纳入200例接受乳腺X线摄影的女性,随机分组在头尾位(CC)对一侧乳腺采用自我压迫,比较压迫力、乳腺厚度和平均腺体剂量(MGD)。结果显示自我压迫使压迫力显著增加21.7%(19.8 N),乳腺厚度减少5%(2.43 mm),MGD降低6.3%(0.09 mGy),图像质量评估无差异,且与自我压迫侧别无关。
要点
- 自我压迫可降低CC位平均腺体剂量约6.3%,同时不降低图像质量。
- 自我压迫使压迫力增加约19.8 N,乳腺厚度减少约2.43 mm。
- 自我压迫效果与实施侧别无关,可考虑在CC位协议中采用。
诊断标准与关键数值
- 自我压迫使压迫力显著增加21.7%(19.8 N)。
- 自我压迫使乳腺厚度显著减少5%(2.43 mm)。
- 自我压迫使平均腺体剂量(MGD)显著降低6.3%(0.09 mGy)。
- 图像质量按欧洲委员会标准评估,自我压迫与标准流程间无差异。
临床应用与阅片要点
- 在CC位乳腺X线摄影中可考虑采用自我压迫以降低腺体剂量并保证图像质量。
- 自我压迫可增加压迫力并减少乳腺厚度,操作时需关注患者耐受性。
局限性
- 研究仅涉及CC投影,未评估其他体位。
- 样本为单中心200例,结果外推需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The purpose of the study was to investigate whether the use of self-compression in craniocaudal (CC) projection has an effect on compression force, breast thickness, radiation dose and image quality compared to the standard mammographic procedure.
METHODS: The study was conducted on 200 female patients that were referred for mammographic imaging. Patients were randomly divided into two equal groups. In the first group, self-compression was performed on the right breast and in the second group on the left breast. The data about compression force (N), breast thickness (mm), and mean glandular dose (MGD; mGy) were collected. In addition, the differences in the mentioned variables according to the side of self-compression were compared. All mammographic images were evaluated by two experienced radiologists according to the criteria established by the European Commission.
RESULTS: The use of self-compression resulted in a significant increase in compression force by 21.7 % (19.8 N) and a significant reduction of breast thickness by 5% (2.43 mm) and MGD by 6.3 % (0.09 mGy), respectively. There were no statistically significant differences based on the self-compression side, and no differences were observed in image quality assessment.
CONCLUSION: This study demonstrates that the imaging protocol in mammography of the CC projection can be adopted by the use of self-compression in order to achieve better results.