利用定量EPR氧分压图改进DCE-MRI校正的18F-MISO PET肿瘤乏氧定位

Improving Tumor Hypoxia Location in 18F-Misonidazole PET with Dynamic Contrast-enhanced MRI Using Quantitative Electron Paramagnetic Resonance Partial Oxygen Pressure Images.
PubMed 文献 · Radiol Imaging Cancer · 2021-03-26

该研究以小鼠乳腺癌模型为对象,用EPR氧分压图为参考标准,评估DCE-MRI信息校正18F-MISO PET乏氧图像的空间准确性。结果显示校正后PET与EPR乏氧区域的重叠度显著提高,DSC由0.35升至0.86,Hausdorff距离由5.70 mm降至2.29 mm(P < .001)。

要点

  • DCE-MRI校正可显著提高18F-MISO PET乏氧区域与EPR pO2参考标准的空间重叠度。
  • 校正后DSC由0.35±0.23升至0.86±0.18,Hausdorff距离由5.70±1.7 mm降至2.29±0.70 mm。
  • 独立测试小鼠中校正效果同样显著,DSC由0.19±0.28升至0.80±0.29(P = .02)。

关键影像表现/征象

  • 18F-MISO PET与EPR pO2图像乏氧区域在原始PET中重叠度较低,DSC为0.35±0.23,Hausdorff距离为5.70 mm±1.7。
  • 经DCE-MRI校正后PET图像与EPR乏氧区域重叠度提高,DSC升至0.86±0.18,Hausdorff距离降至2.29 mm±0.70。
  • 四只独立测试小鼠中校正后DSC由0.19±0.28升至0.80±0.29(P = .02),Hausdorff距离由6.40 mm±2.5降至1.95 mm±0.63(P = .01)。

诊断标准与关键数值

  • PET图像乏氧阈值定义为肿瘤-肌肉比值大于或等于2.2。
  • EPR图像乏氧阈值定义为氧分压小于或等于14 mm Hg。
  • 以Dice相似系数(DSC)和Hausdorff距离(dH)量化PET与EPR乏氧区域的三维重叠程度。

临床应用与阅片要点

  • DCE-MRI信息可作为校正18F-MISO PET乏氧定位的辅助手段,提高乏氧区域空间准确性。
  • 阅片时需注意原始PET乏氧范围可能与真实pO2分布存在空间偏差,校正后可改善重叠度。

局限性

  • 研究为小鼠动物实验,样本量10只,结论外推至临床需进一步验证。
  • DCE-MRI权重系数基于6只训练小鼠计算,样本量较小。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To enhance the spatial accuracy of fluorine 18 (18F) misonidazole (MISO) PET imaging of hypoxia by using dynamic contrast-enhanced (DCE) MR images as a basis for modifying PET images and by using electron paramagnetic resonance (EPR) partial oxygen pressure (pO2) as the reference standard.

MATERIALS AND METHODS: Mice (n = 10) with leg-borne MCa4 mammary carcinomas underwent EPR imaging, T2-weighted and DCE MRI, and 18F-MISO PET/CT. Images were registered to the same space for analysis. The thresholds of hypoxia for PET and EPR images were tumor-to-muscle ratios greater than or equal to 2.2 mm Hg and less than or equal to 14 mm Hg, respectively. The Dice similarity coefficient (DSC) and Hausdorff distance (d H ) were used to quantify the three-dimensional overlap of hypoxia between pO2 EPR and 18F-MISO PET images. A training subset (n = 6) was used to calculate optimal DCE MRI weighting coefficients to relate EPR to the PET signal; the group average weights were then applied to all tumors (from six training mice and four test mice). The DSC and d H were calculated before and after DCE MRI-corrected PET images were obtained to quantify the improvement in overlap with EPR pO2 images for measuring tumor hypoxia.

RESULTS: The means and standard deviations of the DSC and d H between hypoxic regions in original PET and EPR images were 0.35 mm ± 0.23 and 5.70 mm ± 1.7, respectively, for images of all 10 mice. After implementing a preliminary DCE MRI correction to PET data, the DSC increased to 0.86 mm ± 0.18 and the d H decreased to 2.29 mm ± 0.70, showing significant improvement (P < .001) for images of all 10 mice. Specifically, for images of the four independent test mice, the DSC improved with correction from 0.19 ± 0.28 to 0.80 ± 0.29 (P = .02), and the d H improved from 6.40 mm ± 2.5 to 1.95 mm ± 0.63 (P = .01).

CONCLUSION: Using EPR information as a reference standard, DCE MRI information can be used to correct 18F-MISO PET information to more accurately reflect areas of hypoxia.Keywords: Animal Studies, Molecular Imaging, Molecular Imaging-Cancer, PET/CT, MR-Dynamic Contrast Enhanced, MR-Imaging, PET/MR, Breast, Oncology, Tumor Mircoenvironment, Electron Paramagnetic ResonanceSupplemental material is available for this article.© RSNA, 2021.