单中心横断面研究纳入133例慢性肝病患者,评估Aixplorer Mach 30超声系统上2-D-SWE PLUS和SSp PLUS定量肝纤维化与脂肪变,以瞬时弹性成像联合CAP为参考。2-D-SWE PLUS与TE测肝硬度高度相关(r=0.92),SSp PLUS与CAP强负相关(r=-0.70),并给出预测纤维化分期和脂肪变的cutoff值。

要点

  • 2-D-SWE PLUS测肝硬度与TE高度相关(r=0.92),可用于无创纤维化评估。
  • 预测F≥2、F≥3、F4的2-D-SWE PLUS cutoff分别为6.8、8.4、11 kPa。
  • SSp PLUS预测脂肪变的最佳cutoff为1537 m/s,与CAP强负相关(r=-0.70)。
英文原文(摘要)· 点击展开

The goal of this study was to evaluate the performance of two new ultrasound-based techniques (ShearWave PLUS elastography [2-D-SWE PLUS], SSp PLUS Imaging [SSp PLUS]) implemented on the Aixplorer Mach 30 ultrasound system (Supersonic Imagine, Aix-en-Provence, France) for the non-invasive assessment of liver steatosis and fibrosis, using transient elastography (TE) with the controlled attenuation parameter (CAP) as reference. This monocentric cross-sectional study included 133 consecutive adult patients with chronic hepatopathies. Liver stiffness and steatosis were evaluated in the same session using the techniques mentioned above. An excellent correlation was observed between liver stiffness measurements obtained with 2-D-SWE PLUS and TE (r = 0.92, p < 0.0001). The best cutoff value of 2-D-SWE PLUS for predicting F≥2 was 6.8 kPa; for F≥3, 8.4 kPa; and for F4, 11 kPa. With respect to steatosis evaluation, a strong negative correlation between CAP and SSp PLUS values (r = -0.70, p < 0.0001) was obtained. The best SSp PLUS cutoff value for predicting steatosis was 1537 m/s.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Popa A, Șirli R, Popescu A, et al. Ultrasound-Based Quantification of Fibrosis and Steatosis with a New Software Considering Transient Elastography as Reference in Patients with Chronic Liver Diseases. Ultrasound Med Biol. 2021. PMID: 33832824. doi:10.1016/j.ultrasmedbio.2021.02.029

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。