低管电压下冠状动脉钙化评分:剂量确定与评分机制

Coronary artery calcium scoring at lower tube voltages - Dose determination and scoring mechanism.
PubMed 文献 · Eur J Radiol · 2021-04-05

研究利用两种体模和双源CT,建立低管电压下CTDIvol与120 kVp基线CTDIvol的幂函数关系(指数1.246),并通过图像降尺度保持常规Agatston阈值,使低电压钙化评分与真实值偏差在12%以内。

要点

  • 低管电压钙化评分需按幂函数关系确定剂量并用降尺度因子校正图像。
  • 降尺度因子0.88-0.63可保持常规Agatston阈值不变。
  • 校正后低电压钙化评分与真实值偏差在12%以内。

局限性

  • 研究基于体模,未在临床患者中验证。
  • 样本量有限,仅两种体模尺寸。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Population dose has been a concern with coronary artery calcium scoring CT since it is performed in adults with borderline risk. Lower tube voltage acquisitions are appealing but there are no agreed schemes for reduced dose determination. Moreover, conventional scoring cannot be used without changing the multiple Agatston thresholds.

METHODS: By applying consistent calcium contrast-to-noise ratio to two anthropomorphic heart phantoms (medium and large) with 3-cm hydroxyapatite (HA) inserts, scanned using a dual-source CT, the relationship was derived between the volume CT dose index (CTDIvol) at lower tube voltages and the baseline CTDIvol at 120 kVp. The baseline CTDIvol was obtained using the noise thresholds from the images acquired at 120 kVp. To preserve the conventional Agatston thresholds, down-scaling with the found factors was applied to images acquired at lower voltages with a dynamic heart module and 1.2-5 mm inserts (50-400 mg/cc) on the coronary tracks. Scores were evaluated on the scaled images by six readers.

RESULTS: The CTDIvol at lower voltages was related to the baseline CTDIvol following a power form of the voltage (index 1.246), regardless of the phantom size. The baseline CTDIvol was 1.5 and 4.5 mGy, for the medium and large phantoms, respectively. Correspondingly, the reduced CTDIvol at 100-70 kVp were 1.28-0.76 mGy, and 3.57-2.32 mGy. The downscaling factors were 0.88-0.63. The calcium scores at lower voltages were found within 12 % of the ground-truths.

CONCLUSION: A vendor-independent approach was established to obtain the reduced dose and correct coronary calcium scores at lower tube voltages.

来源:Eur J Radiol,查看原文。版权归原作者所有。