文献速递心血管CT

光子计数CT评估28种冠状动脉支架的体外研究

Photon-Counting Computed Tomography for Coronary Stent Imaging: In Vitro Evaluation of 28 Coronary Stents.
PubMed 文献 · Invest Radiol · 2021-01-01

本体外模体研究比较了光子计数探测器CT与能量积分探测器CT对28种冠状动脉支架的支架内管腔可见性和定量图像特征。结果显示,光子计数CT在标准分辨率和高分辨率模式下均显著提高支架内管腔可见性(中位数66.7%和68.9% vs 65.4%),降低支架内CT衰减增量(78 HU和85 HU vs 108 HU)和图像噪声(21 HU vs 25 HU),表明其图像质量优于传统CT。

要点

  • 光子计数CT可提高支架内管腔可见性,减少支架伪影干扰。
  • 高分辨率模式对支架内管腔显示最佳,标准分辨率模式噪声最低。
  • 支架内CT衰减增量较低,有助于减少对管腔评估的高估。

诊断标准与关键数值

  • 支架内管腔可见性中位数:光子计数CT标准分辨率66.7%(IQR 63.3-72.3%),高分辨率68.9%(IQR 64.4-74.4%),能量积分CT 65.4%(IQR 62.2-70.4%),差异有统计学意义(P<0.001)。
  • 支架内CT衰减增量中位数:光子计数CT标准分辨率78 HU(IQR 46-108,P=0.024),高分辨率85 HU(IQR 59-113,P=0.006),能量积分CT 108 HU(IQR 85-126)。
  • 图像噪声中位数:光子计数CT标准分辨率21 HU(IQR 21-21),能量积分CT 25 HU(IQR 24-25.0),差异有统计学意义(P<0.001)。

局限性

  • 本研究为体外模体研究,结果需在临床患者中进一步验证。
  • 仅评估了图像质量参数,未涉及诊断准确性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to assess in-stent lumen visibility and quantitative image characteristics of different coronary stents using a novel photon-counting detector (PCD) computed tomography (CT) system in comparison to a state-of-the-art energy-integrating detector (EID) CT scanner.

MATERIALS AND METHODS: In this in vitro phantom study, 28 different coronary stents ranging from 2.25 to 4.5 mm lumen diameter were expanded into plastic tubes filled with contrast agent. Stent-containing plastic tubes were positioned in a custom-made emulsion-filled phantom, which was inserted into an anthropomorphic phantom simulating a medium-sized patient. Computed tomography scans were acquired parallel to the scanners' z axis using a novel cadmium telluride-based PCD CT system (SOMATOM CountPlus; Siemens Healthcare GmbH, Forchheim Germany), operating in 2 different modes (standard-resolution mode [SR] and ultra-high-resolution [UHR] mode), and a latest generation dual-source EID CT system (SOMATOM Force; Siemens Healthcare GmbH, Forchheim). CTDIvol-matched images were reconstructed with comparable convolution kernels and using the same reconstruction parameters. In-stent lumen visibility (in %), increase in in-stent attenuation (expressed as Δ in-stent CT attenuation), and image noise (in Hounsfield unit) were manually measured. Parts of the image analysis (in-stent lumen visibility) were additionally performed in an automated way. Differences were tested using Wilcoxon signed rank test.

RESULTS: The best in-stent lumen visibility was achieved with the PCD-UHR mode and the lowest noise levels with the PCD-SR mode. The median in-stent lumen visibility was significantly higher (P < 0.001) with PCD (SR, 66.7%; interquartile range [IQR], 63.3-72.3; UHR, 68.9%; IQR, 64.4-74.4) compared with EID (65.4%; IQR, 62.2-70.4). The Δ in-stent CT attenuation was significantly lower for PCD in both SR (78 HU; IQR, 46-108; P = 0.024) and UHR (85 HU; IQR, 59-113; P = 0.006) compared with EID (108 HU; IQR, 85-126). Image noise was significantly lower (P < 0.001) for PCD-SR (21 HU; IQR, 21-21) compared with EID images (25 HU; IQR, 24-25.0).

CONCLUSIONS: The PCD provides superior in-stent lumen visibility and quantitative image characteristics when compared with conventional EID.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Petritsch B, Petri N, Weng AM, et al. Photon-Counting Computed Tomography for Coronary Stent Imaging: In Vitro Evaluation of 28 Coronary Stents. Invest Radiol. 2021. PMID: 33867450. doi:10.1097/RLI.0000000000000787

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。