文献速递胸部CT

宽体探测器CT宽容积扫描模式在间质性肺疾病中降低辐射剂量的研究

Radiation dose reduction with the wide-volume scan mode for interstitial lung diseases.
PubMed 文献 · Eur Radiol · 2021-04-15

这项回顾性单中心研究纳入50例间质性肺疾病随访或筛查患者,对比宽容积模式与常规螺旋CT的辐射剂量和图像质量。宽容积模式中位CTDIvol和DLP分别为3.1 mGy和94.6 mGy·cm,较螺旋模式(3.7 mGy和122.1 mGy·cm)显著降低,相对降低21%和32%,图像噪声和质量无显著差异。结论认为宽容积模式可常规用于间质性肺疾病的诊断和随访。

要点

  • 宽容积模式较螺旋模式降低CTDIvol约21%、DLP约32%,图像质量无显著差异。
  • 宽容积模式偶尔出现容积交界处错位伪影,但多数不影响诊断质量。
  • 间质性肺疾病俯卧位扫描可优先考虑宽容积模式以减少辐射剂量。

关键影像表现/征象

  • 宽容积模式与螺旋模式的图像噪声和整体图像质量无显著差异。
  • 宽容积扫描偶尔可见两个容积交界处的错位伪影,但多数病例不损害诊断质量。
  • 研究分析了正常结构、病变、整体图像质量和伪影,未报告具体影像征象细节。

诊断标准与关键数值

  • 宽容积模式中位CTDIvol为3.1 mGy,螺旋模式为3.7 mGy,p<0.0001。
  • 宽容积模式中位DLP为94.6 mGy·cm,螺旋模式为122.1 mGy·cm,p<0.0001。
  • CTDIvol和DLP的相对降低分别为21%和32%。

临床应用与阅片要点

  • 间质性肺疾病随访或筛查时,俯卧位补充扫描可采用宽容积模式以降低辐射剂量。
  • 阅片时注意宽容积扫描容积交界处的错位伪影,多数不影响诊断。
  • 根据胸廓长度选择320×0.5mm或240×0.5mm准直。

局限性

  • 回顾性单中心研究,样本量50例。
  • 宽容积模式仅用于俯卧位补充扫描,未与仰卧位螺旋扫描直接同体位对比。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The wide-volume mode, available on wide-area detector row CTs, has the advantage of reducing exposure time and radiation dose. It is infrequently used for lung diseases. The purpose of this study is to compare image quality and radiation dose of wide-volume chest CT to those of standard helical CT in the setting of interstitial lung diseases.

METHODS: Retrospective monocentric study including 50 consecutive patients referred for follow-up or screening of interstitial lung diseases, requiring prone scan, acquired with the wide-volume mode, in addition to the routine supine scan, acquired with the helical mode. The optimal collimation in wide-volume mode (320 × 0.5mm or 240 × 0.5mm) was chosen according to the length of the thorax. Wide-volume acquisitions were compared to helical acquisitions for radiation dose (CTDIvol, DLP) and image quality, including analysis of normal structures, lesions, overall image quality, and artifacts (Wilcoxon signed-rank test).

RESULTS: Median CTDIvol and DLP with wide volumes (3.1 mGy and 94.6 mGy·cm) were significantly reduced (p < 0.0001) as compared to helical mode (3.7mGy and 122.1 mGy·cm), leading to a median 21% and 32% relative reduction of CTDIvol and DLP, respectively. Image noise and quality were not significantly different between the two modes. Misalignment artifact at the junction of two volumes was occasionally seen in the wide-volume scans and, when present, did not impair the diagnostic quality in the majority of cases.

CONCLUSIONS: Wide-volume mode allows 32% radiation dose reduction compared to the standard helical mode and could be used routinely for diagnosis and follow-up of interstitial lung diseases.

KEY POINTS: • Retrospective monocentric study showed that wide-volume scan mode reduces radiation dose by 32% in comparison to helical mode for chest CT in the setting of interstitial lung diseases. • Mild misalignment may be observed at the junction between volumes with the wide-volume mode, without decrease of image quality in the majority of cases and without impairing diagnostic quality. • Wide-volume mode could be used routinely for the diagnosis and follow-up of interstitial lung diseases.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Abdo A, Karam E, Henry T, et al. Radiation dose reduction with the wide-volume scan mode for interstitial lung diseases. Eur Radiol. 2021. PMID: 33856516. doi:10.1007/s00330-021-07862-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。