针内超声换能器用于肋间神经阻滞:一项概念验证研究
本研究设计了一种针内超声(INUS)系统,换能器为20 MHz单晶,可置入18G针,距针尖1 mm处发射射频模式超声信号。在10只仔猪中完成100次肋间穿刺,所有穿刺均成功识别肋间肌、胸内筋膜及双层胸膜,证实INUS可实时识别肋间结构并辅助肋间神经阻滞。
要点
- INUS换能器可置入18G针,距针尖1 mm发射20 MHz射频超声信号。
- 100次肋间穿刺均成功识别肋间肌、胸内筋膜和双层胸膜。
- 胸膜具有特征性超声信号,有助于判断针尖位置和阻滞成功。
核心结论
- INUS可实现肋间结构的实时识别,并有助于成功实施肋间神经阻滞。
研究设计与样本
- 研究为概念验证研究,在10只仔猪中完成100次肋间穿刺。
- INUS换能器为20 MHz单晶,可置入常规18G针,增益设为20 dB,距针尖1 mm处发射射频模式超声信号。
- 肋间间隙通过表面超声或触诊识别,并插入INUS换能器针推进直至识别目标解剖结构。
- 阻滞成功定义为理想的生理盐水和染料扩散,并通过解剖确认。
关键影像表现/征象
- 胸膜具有特征性超声信号。
诊断标准与关键数值
- 阻滞成功定义为理想的生理盐水和染料扩散,并通过解剖确认。
- 成功检测肋间肌、胸内筋膜及双层脏层和壁层胸膜在全部100次穿刺尝试中均实现。
临床应用与阅片要点
- INUS可在肋间神经阻滞过程中实时识别肋间肌、胸内筋膜和双层胸膜,辅助针尖定位。
- 胸膜的特征性超声信号可作为判断针尖位置和阻滞成功的参考。
局限性
- 研究为动物实验(仔猪),样本量有限(10只仔猪、100次穿刺),结论向临床转化需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Intercostal nerve block is a widely used and effective approach to providing regional anesthesia in the thoracic region for pain relief. However, during ultrasound-guided intercostal nerve block, inaccurate identification of the anatomic structures or suboptimal positioning of the needle tip may result in complications and blockade failure. In this study, we designed an intraneedle ultrasound (INUS) system and validated its efficacy in identifying anatomic structures relevant to thoracic region anesthesia. The 20-MHz INUS transducer comprised a single lead magnesium niobate-lead titanate crystal, and gain was set to 20 dB. It fit into a regular 18G needle and emitted radiofrequency-mode ultrasound signals at 1 mm from the needle tip. One hundred intercostal punctures were performed in 10 piglets. Intercostal spaces were identified by surface ultrasound or palpation and located by inserting and advancing the INUS transducer needle until the appropriate anatomy was identified. Blockade success was defined by ideal saline and dye spreading and confirmed by dissection. The pleura had a distinctive ultrasound signal, and successful detection of the intercostal muscles, endothoracic fascia and double-layered parietal and visceral pleura was achieved in all 100 puncture attempts. INUS allows real-time identification of intercostal structures and facilitates successful intercostal nerve blocks.
引用
- Liao SW, Su FW, Ting CK, et al. Intercostal Nerve Block Using an Innovative Intraneedle Ultrasound Transducer: A Proof-of-Concept study. Ultrasound Med Biol. 2021. PMID: 33836903. doi:10.1016/j.ultrasmedbio.2021.03.001
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。