Inhance 3D Velocity非增强MR静脉成像诊断颅内静脉血栓及静脉窦发育不良的影响

Non-contrast magnetic resonance venography with Inhance 3D Velocity: diagnostic performance for intracranial venous thrombosis.
PubMed 文献 · Neuroradiology · 2021-04-06

本研究纳入540例患者,以对比增强MR静脉成像联合常规序列为金标准,评估Inhance 3D Velocity(I3DV)诊断颅内静脉血栓的性能。结果显示I3DV诊断敏感度95.7%、特异度92.1%、准确率92.6%,且假阳性血栓诊断与左横窦发育不良显著相关。

要点

  • I3DV诊断颅内静脉血栓敏感度95.7%、特异度92.1%,可用于对比剂禁忌者。
  • 左横窦发育不良在I3DV上易被误判为血栓,需注意鉴别。
  • I3DV能区分完全性与部分性血栓,与金标准一致性显著。

局限性

  • 单中心研究,可能存在选择偏倚。
  • 未报告阅片者间一致性及设备参数细节。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The aim of this study was to evaluate the diagnostic performance of Inhance 3D Velocity (I3DV) in intracranial venous thrombosis and investigate the possible impact of venous sinus hypoplasia/aplasia on false thrombosis diagnosis made with I3DV.

METHODS: This study included 540 patients. Contrast-enhanced magnetic resonance venography combined with conventional sequences was considered the gold standard test (GST), while I3DV was considered as diagnostic test. We accessed the diagnostic success of I3DV for intracranial venous thrombosis detection, thrombosed vessel identification, and total/partial thrombus distinction. The possible relationship between false-positive thrombus diagnosed by I3DV and venous sinus hypoplasia or aplasia diagnosed by GST was investigated.

RESULTS: The sensitivity, specificity, positive predictive value, negative predictive value, and accuracy of I3DV in the detection of intracranial venous thrombosis were 95.7%, 92.1%, 64.1%, 99.3%, and 92.6%, respectively. A significant association was observed between I3DV and GST in terms of thrombosis detection and total/partial thrombus distinction (p < 0.001). A significant relationship was observed between false-positive thrombosis diagnosis in I3DV and hypoplasia in the left transverse sinus (p < 0.001).

CONCLUSION: Intracranial venous thrombosis may be diagnosed faster and more accurately than traditional phase contrast magnetic resonance angiography in I3DV. This technique can be used in situations where contrast medium application is contraindicated. As in other non-contrast magnetic resonance venography techniques, left transverse sinus hypoplasia can be diagnosed as a thrombosed vessel in I3DV.

来源:Neuroradiology,查看原文。版权归原作者所有。