压缩敏感度编码加速脑TOF-MRA的多中心临床研究
结果显示CS4在图像质量与采集时间间平衡最佳,推荐用于常规脑TOF-MRA;CS6适用于不评估小动脉时;CS10用于大动脉快速可视化。
要点
- CS4推荐用于常规脑TOF-MRA,兼顾图像质量与采集时间。
- CS6可用于不评估小动脉的检查,因2根小动脉显示不可靠。
- CS10仅适用于大动脉快速可视化,大动脉显示可接受。
关键影像表现/征象
- 随着CS加速因子增加,所评估动脉的重建信号强度和结构相似性持续下降。
- SF2、CS2和CS4两两比较,对比率无显著差异。
- RS、SF2和CS2两两比较,定性评估无显著差异(P > .05)。
- CS2和CS4图像中所有评估动脉显示可接受;CS6图像中2根小动脉显示不可靠;CS8和CS10图像中大动脉显示可接受。
诊断标准与关键数值
- 定量评估指标包括重建信号强度、结构相似性和对比率。
- 定性评估包括动脉评分、伪影、整体图像质量和动脉瘤及狭窄的诊断信心评分。
- 评估动脉包括5根大动脉、6根中动脉和6根小动脉。
临床应用与阅片要点
- 常规脑TOF-MRA可优先选择CS4以平衡图像质量与采集时间。
- 若检查目的不涉及小动脉评估,可考虑CS6以进一步缩短采集时间。
- 仅需快速显示大动脉时,可选择CS10。
- 注意CS6可能对小动脉显示不可靠,CS8和CS10仅大动脉显示可接受。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: The clinical practice of three-dimensional TOF-MRA, despite its capability in brain artery assessment, has been hampered by the relatively long scan time, while recent developments in fast imaging techniques with random undersampling has shed light on an improved balance between image quality and imaging speed. Our aim was to evaluate the effectiveness of TOF-MRA accelerated by compressed sensitivity encoding and to identify the optimal acceleration factors for routine clinical use.
MATERIALS AND METHODS: One hundred subjects, enrolled at 5 centers, underwent 8 brain TOF-MRA sequences: 5 sequences using compressed sensitivity encoding with acceleration factors of 2, 4, 6, 8, and 10 (CS2, CS4, CS6, CS8, and CS10), 2 using sensitivity encoding with factors of 2 and 4 (SF2 and SF4), and 1 without acceleration as a reference sequence (RS). Five large arteries, 6 medium arteries, and 6 small arteries were evaluated quantitatively (reconstructed signal intensity, structural similarity, contrast ratio) and qualitatively (scores on arteries, artifacts, overall image quality, and diagnostic confidence for aneurysm and stenosis). Comparisons were performed among the 8 sequences.
RESULTS: The quantitative measurements showed that the reconstructed signal intensities of the assessed arteries and the structural similarity consistently decreased as the compressed sensitivity encoding acceleration factor increased, and no significant difference was found for the contrast ratios in pair-wise comparisons among SF2, CS2, and CS4. Qualitative evaluations showed no significant difference in pair-wise comparisons among RS, SF2, and CS2 (P > .05). The visualization of all the assessed arteries was acceptable for CS2 and CS4, while 2 small arteries in images of CS6 were not reliably displayed, and the visualization of large arteries was acceptable in images of CS8 and CS10.
CONCLUSIONS: CS4 is recommended for routine brain TOF-MRA with balanced image quality and acquisition time; CS6, for examinations when small arteries are not evaluated; and CS10, for fast visualization of large arteries.
引用
- Ding J, Duan Y, Zhuo Z, et al. Acceleration of Brain TOF-MRA with Compressed Sensitivity Encoding: A Multicenter Clinical Study. AJNR Am J Neuroradiol. 2021. PMID: 33858820. doi:10.3174/ajnr.A7091
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。