稀疏采样低剂量MDCT在脊柱内固定患者中减少金属伪影的研究
结果显示使用25%投影数据时,所有并发症均被检出,诊断信心高(平均分1.2-1.3),且容积CT剂量指数从12.6 mGy降至3.2 mGy。结论认为稀疏采样联合SIR可大幅降低辐射剂量,25%投影数据足以准确诊断主要内固定相关并发症。
要点
- 25%投影数据联合SIR未漏诊任何内固定相关并发症。
- 25%投影数据下诊断信心仍高(平均分≤1.3,1分为高信心)。
- 容积CT剂量指数从12.6 mGy降至3.2 mGy。
核心结论
- 稀疏采样联合SIR用于脊柱内固定患者低剂量MDCT可显著降低辐射暴露,25%投影数据未漏诊并发症且诊断信心高。
研究设计与样本
- 两名阅片者定性评估整体图像质量、伪影、图像对比、椎管观察及诊断信心(1=高,2=中,3=低)。
关键影像表现/征象
- 随着投影数减少,整体图像质量下降、伪影增加。
- 使用100%、50%、25%投影数据时,两名阅片者均检出所有并发症。
- 25%投影数据下诊断信心仍高(阅片者1平均分1.2±0.4,阅片者2为1.3±0.5),阅片者间一致性为中等至几乎完美(加权Cohen κ=0.787-0.855)。
诊断标准与关键数值
- 25%投影数据对应容积CT剂量指数为3.2 mGy,而原始采集(100%投影)为12.6 mGy。
- 诊断信心评分标准:1=高信心,2=中等信心,3=低信心。
临床应用与阅片要点
- 在日常阅片中,对于脊柱内固定患者的低剂量MDCT,使用25%投影数据联合SIR可能足以准确检出主要内固定相关并发症。
- 需注意随着投影数减少图像质量下降和伪影增加,但25%投影数据下诊断信心仍高且未漏诊并发症。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The purpose of our study was to evaluate simulated sparse-sampled MDCT combined with statistical iterative reconstruction (SIR) for low-dose imaging of patients with spinal instrumentation. MATERIALS AND METHODS. Thirty-eight patients with implanted hardware after spinal instrumentation (24 patients with short- or long-term instrumentation-related complications [i.e., adjacent segment disease, screw loosening or implant failure, or postoperative hematoma or seroma] and 14 control subjects with no complications) underwent MDCT. Scans were simulated as if they were performed with 50% (P50), 25% (P25), 10% (P10), and 5% (P5) of the projections of the original acquisition using an in-house-developed SIR algorithm for advanced image reconstructions. Two readers performed qualitative image evaluations of overall image quality and artifacts, image contrast, inspection of the spinal canal, and diagnostic confidence (1 = high, 2 = medium, and 3 = low confidence). RESULTS. Although overall image quality decreased and artifacts increased with reductions in the number of projections, all complications were detected by both readers when 100% of the projections of the original acquisition (P100), P50, and P25 imaging data were used. For P25 data, diagnostic confidence was still high (mean score ± SD: reader 1, 1.2 ± 0.4; reader 2, 1.3 ± 0.5), and interreader agreement was substantial to almost perfect (weighted Cohen κ = 0.787-0.855). The mean volumetric CT dose index was 3.2 mGy for P25 data in comparison with 12.6 mGy for the original acquisition (P100 data). CONCLUSION. The use of sparse sampling and SIR for low-dose MDCT in patients with spinal instrumentation facilitated considerable reductions in radiation exposure. The use of P25 data with SIR resulted in no missed complications related to spinal instrumentation and allowed high diagnostic confidence, so using only 25% of the projections is probably enough for accurate and confident diagnostic detection of major instrumentation-related complications.
引用
- Sollmann N, Mei K, Riederer I, et al. Low-Dose MDCT of Patients With Spinal Instrumentation Using Sparse Sampling: Impact on Metal Artifacts. AJR Am J Roentgenol. 2021. PMID: 33703925. doi:10.2214/AJR.20.23083
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。