乳腺剪切波弹性成像:技术图像质量参数对诊断准确性的影响
本回顾性研究纳入281个乳腺病灶,由3名不知病理结果的阅片者按5项参数对SWE图像质量评分(低质量0分、高质量1分),比较Emax、Emean、ESD在高低质量图像上诊断恶性的ROC表现。高质量图像上Emax(p<.001)、Emean(p=.002)、ESD(p<.001)的AUC均显著高于低质量图像。结论提示质量参数可辅助医师判读乳腺SWE图像,并可能提高其鉴别良恶性的准确性。
要点
- B模式病灶显示与近场红 pattern 均评为1分者为高质量图像(n=167),任一为0分者为低质量(n=114)。
- 高质量图像上Emax、Emean、ESD诊断恶性的AUC均显著高于低质量图像。
- 5项质量参数中仅B模式病灶显示、近场红 pattern、周围组织表现3项阅片者间一致性为中等至较高。
诊断标准与关键数值
- 图像质量按5项参数逐项评分:低质量记0分、高质量记1分。
- 5项参数为:双幅显示中B模式病灶可见性、FOV近场SWE红色 pattern(代表高硬度)、周围组织表现、FOV放置、Emax/Emin/Emean/ESD的ROI放置。
- B模式病灶可见性与近场红 pattern 共识评分均为1分者为高质量图像(n=167),任一为0分者为低质量图像(n=114)。
- 以Emax为恶性分类器时高质量图像AUC显著高于低质量图像(p<.001)。
- 以Emean(p=.002)和ESD(p<.001)为恶性分类器时高质量图像AUC亦显著高于低质量图像。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The purpose of this study is to determine the impact of shear-wave elastography (SWE) image quality parameters on the diagnostic performance of elasticity measurements in classifying breast lesions. MATERIALS AND METHODS. This retrospective study included 281 breast lesions that underwent SWE and ultrasound-guided biopsy performed between October 1, 2017, and August 31, 2018. Three readers who were blinded to pathologic outcomes independently scored the image quality of each SWE image (with low quality denoted by a score of 0 and high quality indicated by a score of 1) on the basis of five parameters: B-mode visualization of the lesion on a dual-panel display, SWE red pattern (denoting high stiffness) in the near field of the FOV, appearance of the surrounding tissue, FOV placement, and ROI placement for the maximum (Emax), minimum (Emin), mean (Emean), and SD (ESD) of Young modulus elasticity measurements. Using ROC analysis, we compared the performance of Emax, Emean, and ESD in diagnosing malignancy on low- and high-quality images on the basis of consensus (i.e., majority) scores for each individual quality parameter as well as two models combining a few of the quality parameters. RESULTS. Three quality parameters (B-mode visualization of the lesion, presence of a near-field red pattern, and the appearance of the surrounding tissue) showed moderate-to-substantial interobserver agreement. SWE images were considered high quality (n = 167) if both B-mode visualization and near-field red pattern received a consensus score of 1, and they were considered low quality (n = 114) if either parameter received a consensus score of 0. High-quality images had a statistically higher AUC value than low-quality images when Emax (p < .001), Emean (p = .002), and ESD (p < .001) were used as classifiers of malignancy. CONCLUSION. Quality parameters can support radiologists who are performing and interpreting breast SWE images. These quality parameters have the potential to improve the accuracy of SWE in differentiating malignant from benign breast lesions.
引用
- Chou SS, Baikpour M, Zhang W, et al. Shear-Wave Elastography of the Breast: Impact of Technical Image Quality Parameters on Diagnostic Accuracy. AJR Am J Roentgenol. 2021. PMID: 33729888. doi:10.2214/AJR.19.22728
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。