相变全氟戊烷纳米乳增强脉冲HIFU消融离体灌注肝及肿瘤模型研究

Phase-shift Perfluoropentane Nanoemulsions Enhance Pulsed High-intensity Focused Ultrasound Ablation in an Isolated Perfused Liver System and Their Potential Value for Cancer Therapy.
PubMed 文献 · J Ultrasound Med · 2021-03-16

研究将PLGA包裹的全氟戊烷(PLGA-PFP)纳米乳加入离体灌注肝系统及皮下VX2肿瘤兔模型,观察其对脉冲HIFU消融的增强作用。结果显示PLGA-PFP较PBS对照显著增大HIFU所致坏死体积(P<.05),10%占空比脉冲HIFU的惯性空化强于连续波HIFU(P<.01),50%占空比10s暴露的平均坏死体积显著大于100%占空比(P<.005),而10%占空比则较低(P<.05);兔VX2肿瘤中10%占空比HIFU后平均坏死体积较100%占空比显著增大(P<.05)。

要点

  • PLGA-PFP纳米乳可增强HIFU在离体灌注肝和兔VX2肿瘤中的消融效果,坏死体积较PBS对照显著增大(P<.05)。
  • 脉冲HIFU(10%占空比)的惯性空化显著强于连续波HIFU(P<.01),提示占空比影响空化效应。
  • 离体肝中50%占空比10s暴露的平均坏死体积显著大于100%占空比(P<.005),但10%占空比则较低(P<.05)。

关键影像表现/征象

  • 与PBS对照相比,PLGA-PFP纳米乳显著增强离体灌注肝和皮下VX2肿瘤兔中HIFU诱导的坏死体积(P<.05)。
  • 10%占空比脉冲HIFU暴露的惯性空化远强于连续波HIFU暴露(P<.01)。
  • 100%占空比时的峰值温度显著高于10%占空比(P<.05)。
  • 离体灌注肝中,50%占空比10s暴露的平均坏死体积显著大于100%占空比(P<.005),而10%占空比则较低(P<.05)。
  • 皮下VX2肿瘤兔中,10%占空比HIFU暴露后的平均坏死体积较100%占空比显著增大(P<.05);组织病理学显示脉冲HIFU后为液化性坏死。

临床应用与阅片要点

  • PLGA-PFP纳米乳联合脉冲HIFU可能增大坏死体积,阅片时注意对比增强前后坏死范围的变化。
  • 脉冲HIFU(如10%占空比)惯性空化更强,但峰值温度低于100%占空比,需结合温度监控评估消融效果。
  • 组织病理学显示脉冲HIFU后为液化性坏死,影像评估时需注意与凝固性坏死鉴别。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To investigate whether phase-shift perfluoropetane (PFP) nanoemulsions can enhance pulsed high-intensity focused ultrasound (HIFU) ablation.

METHODS: PFP was encapsulated by poly(lactic-co-glycolic acid) (PLGA) to form a nanometer-sized droplet (PLGA-PFP), which was added to an isolated perfused liver system. Meanwhile, phosphate-buffered saline (PBS) was used as a control. The perfused liver was exposed to HIFU (150 W, t = 3/5/10 s) at various duty cycles (DCs). The ultrasound images, cavitation emissions, and temperature were recorded. Rabbits with subcutaneous VX2 tumors were exposed to HIFU (150 W) at various DCs with or without PLGA-PFP. After ablation, necrosis volume and energy efficiency factor were calculated. Pathologic characteristics were observed.

RESULTS: Compared to the PBS control, PLGA-PFP nanoemulsions markedly enhanced HIFU-induced necrosis volume in both perfused livers and subcutaneous VX2 tumor-bearing rabbits (P <.05). Inertial cavitation was much stronger in the pulsed-HIFU exposure at 10% than that in the continuous-wave HIFU exposure (P <.01). Peak temperature at 100% DC was significantly higher than that at 10% (P <.05). Compared to 100% DC HIFU exposure, the mean necrosis volume induced by 10 s exposure at 50% DC was significantly larger (P <.005) but lower at 10% DC in the isolated perfused livers (P <.05). In addition, the mean necrosis volume in subcutaneous VX2 tumor-bearing rabbits was significantly increased after HIFU exposure at 10% DC when compared to those at 100% DC (P <.05). Histopathologic analysis showed liquefaction necrosis in pulsed HIFU.

CONCLUSION: PLGA-PFP nanoemulsions can enhance HIFU ablation in the isolated perfused livers and promote tumor ablation in the subcutaneous xenograft rabbit model. Appropriate pulsed HIFU exposure may increase the necrosis volume and reduce total ultrasound energy required for HIFU ablation.

来源:J Ultrasound Med,查看原文。版权归原作者所有。